03.04.2013 Views

Gamma3 Long Nail R2.0 Operative Technique - Stryker

Gamma3 Long Nail R2.0 Operative Technique - Stryker

Gamma3 Long Nail R2.0 Operative Technique - Stryker

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong> <strong>R2.0</strong><br />

<strong>Operative</strong> <strong>Technique</strong><br />

Hip Fracture


<strong>Long</strong> <strong>Nail</strong> R .0<br />

Contributing Surgeons:<br />

Prof. Kwok Sui Leung, M. D.<br />

Chairman of Department of Orthopaedics and Traumatology<br />

The Chinese University of Hong Kong<br />

Prince of Wales Hospital<br />

Hong Kong<br />

Dr. Gilbert Taglang<br />

Head of the Trauma Department<br />

Center for Traumatology, Strasbourg<br />

France<br />

Prof. Dr. med. Volker Bühren<br />

Chief of Surgical Services<br />

Medical Director of Murnau Trauma Center, Murnau<br />

Germany<br />

Katsumi Sato M.D.Ph.D.<br />

Vice-Director, Chief Surgeon<br />

Tohoku University Graduate School of Medicine<br />

Tohoku Rosai Hospital, Sendai<br />

Japan<br />

Christopher T. Born M.D.<br />

Professor of Orthopaedic Surgery<br />

Temple University<br />

Philadelphia, PA<br />

USA<br />

Robert Probe, M.D.<br />

Division of Orthopaedic Surgery<br />

Scott & White Memorial Hospital, Temple, Tx<br />

USA<br />

Prof. Dr. med. Vilmos Vécsei<br />

Chief of Traumatology Department<br />

University of Vienna, Vienna<br />

Austria<br />

This publication sets forth detailed<br />

recommended procedures for<br />

using <strong>Stryker</strong> Trauma devices and<br />

instruments.<br />

It offers guidance that you should<br />

heed, but, as with any such technical<br />

guide, each surgeon must consider the<br />

particular needs of each patient and<br />

make appropriate adjustments when<br />

and as required. A workshop training<br />

is required prior to first surgery.<br />

Note:<br />

All bone screws referenced in this<br />

material here are not approved for<br />

screw attachment or fixation to the<br />

posterior elements (pedicles) of the<br />

cervical, thoracic or lumbar spine.


Contents<br />

Introduction<br />

Design Features of the <strong>Gamma3</strong> System<br />

Lag Screw and Set Screw Function<br />

Distal Locking Screws<br />

<strong>Gamma3</strong> System Benefits<br />

Indications/Contraindications<br />

<strong>Operative</strong> <strong>Technique</strong><br />

Implant Selection<br />

Patient Positioning and Fracture Reduction<br />

Special <strong>Technique</strong>s for Fracture Reduction<br />

Incision<br />

Entry Point<br />

Preparation of Medullary Canal<br />

One Step Conical Reamer<br />

Cannulated Cutter<br />

Assembly of Targeting Device<br />

<strong>Nail</strong> Insertion and Positioning<br />

Lag Screw Positioning using One Shot Device<br />

Lag Screw Insertion<br />

Lag Screw Fixation<br />

Distal Screw Locking<br />

End Cap Insertion<br />

<strong>Nail</strong> Extension End Caps<br />

Postoperative Care and Rehabilitation<br />

Extraction of the <strong>Gamma3</strong> Implant<br />

Dealing with Special Cases<br />

Ordering Information – Implants<br />

Ordering Information – Instruments<br />

Publications<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

13<br />

14<br />

16<br />

16<br />

18<br />

20<br />

21<br />

24<br />

25<br />

26<br />

31<br />

34<br />

38<br />

39<br />

39<br />

40<br />

42<br />

43<br />

48<br />

51


Introduction<br />

Introduction<br />

The <strong>Gamma3</strong> Locking <strong>Nail</strong> System is<br />

based on more than 16 years of Gamma<br />

<strong>Nail</strong> experience. This is the third<br />

generation of intramedullary short<br />

and long Gamma fixation nails.<br />

The Evolution of the successful Trochanteric<br />

and <strong>Long</strong> Gamma <strong>Nail</strong>s as<br />

well as the Asia Pacific and Japanese<br />

versions followed strictly a step by<br />

step improvement based on the clinical<br />

experience of the clinical outcome<br />

from surgeons all over the world.<br />

The new <strong>Gamma3</strong> System is designed<br />

to facilitate minimally invasive surgery<br />

and reduce the OR time down to<br />

a minimum by the aid of using new<br />

instrumentation and an optimized<br />

surgical technique.<br />

The nails have a proximal diameter of<br />

15.5mm to help minimize the incision<br />

length required for minimally invasive<br />

surgery. Nevertheless, they offer<br />

the same biomechanical strength and<br />

cut-out resistance as the well established<br />

Trochanteric and <strong>Long</strong> Gamma<br />

<strong>Nail</strong>s.<br />

The new Lag Screw shape has been<br />

improved, especially in the area of the<br />

thread and the cutting flutes at the tip<br />

of the screw. The new design offers<br />

superior cutting behavior during Lag<br />

Screw insertion, providing extremely<br />

low insertion torque. The new thread<br />

design also offers excellent grip in the<br />

cancellous bone of the femoral head<br />

and strong resistance against cut-out.<br />

The 5mm distal locking screws are<br />

currently used in the Gamma-Ti and<br />

the T2 intramedullary nailing systems.<br />

A major advantage of the system is the<br />

newly designed instrument platform.<br />

The instruments are designed for a<br />

minimally invasive surgical technique<br />

and reduce OR time to a minimum.<br />

The instruments are easy to use and<br />

easy to clean, and they share the same<br />

platform as the <strong>Stryker</strong> intramedullary<br />

T2 and S2 nails.<br />

Acknowledgements:<br />

Our thanks are due to the many surgeons<br />

who supported the development<br />

of the new <strong>Gamma3</strong> System, with their<br />

feedback and ideas, during worldwide<br />

panel meetings and helped the<br />

<strong>Gamma3</strong> System to be what it is today.<br />

Special thanks to the Asian Pacific<br />

Technical Committee, who supported<br />

very early the idea of smaller implants<br />

for the treatment of proximal femur<br />

fractures.


Features<br />

Design Features of the <strong>Gamma3</strong><br />

System<br />

<strong>Gamma3</strong> Locking <strong>Nail</strong>s come in 3<br />

neck-shaft angles of 120, 125 and 130°.<br />

• In the following, these <strong>Gamma3</strong><br />

<strong>Nail</strong>s are called:<br />

<strong>Long</strong> <strong>Nail</strong><br />

All nails* use the same Lag Screws, Set<br />

Screw, distal Locking Screws and End<br />

Caps (see Fig. 1).<br />

<strong>Gamma3</strong> <strong>Nail</strong> <strong>Long</strong><br />

This nail incorporates several important<br />

mechanical design features. The<br />

nail is unslotted and cannulated for<br />

Guide-Wire-controlled insertion. To<br />

facilitate conformity with the human<br />

anatomy, the <strong>Long</strong> <strong>Nail</strong> is supplied in a<br />

left and right version.<br />

The three neck-shaft angles accommodate<br />

variations in femoral neck<br />

anatomy. The <strong>Long</strong> <strong>Nail</strong> offers the<br />

opportunity to use two distal Locking<br />

Screws that are inserted through the<br />

distal nail end to control rotation and<br />

telescoping. As shown below, the nail<br />

offers the possibility for either static,<br />

dynamic or secondary dynamic distal<br />

locking, depending on the fracture<br />

pattern.<br />

<strong>Long</strong> <strong>Nail</strong> Distal Locking Options<br />

* Each nail is supplied sterile packaged together with a Set Screw in one box.<br />

• Material:<br />

Titanium alloy with anodized type<br />

II surface treatment or Orthinox®<br />

High Strength Stainless Steel<br />

• <strong>Nail</strong> length:<br />

280mm to 460mm, in 20mm<br />

increments, shorter or longer nails<br />

are available on request<br />

• <strong>Nail</strong> diameter:<br />

proximal 15.5mm, distal: 11.0mm<br />

• Proximal <strong>Nail</strong> angle range:<br />

120°, 125°, 130°<br />

• M-L bend for valgus curvature:<br />

4 degrees<br />

• Proximal anterversion of 10°<br />

• End Caps<br />

0mm, +5mm and +10mm<br />

• Antecurvature radius R .0m of<br />

the shaft<br />

• Distal locking holes<br />

(round and oblong)<br />

for 5mm screws; up to 5mm<br />

dynamization is possible<br />

• Locking in the distal part of the oblong hole creates a dynamic locking<br />

mechanism − requires only one screw (see Fig. 2).<br />

• One screw placed in the distal part of the oblong hole and the other in the<br />

round hole.<br />

If dynamization is required after a period of time, the screw, placed in the<br />

round hole, has to be removed − requires two screws (see Fig. 3).<br />

• One screw placed in the round hole and the other is placed in the proximal<br />

part of the oblong hole − requires two screws (see Fig. 4).<br />

Fig. 2<br />

Dynamic Locking<br />

Fig. 3<br />

Secondary Dynamization<br />

Technical Specifications:<br />

Remove this screw to<br />

allow for dynamization<br />

Fig. 4<br />

Static Locking<br />

Distal Locking Screws<br />

<strong>Gamma3</strong> End Cap<br />

<strong>Gamma3</strong> Set Screw<br />

120°<br />

125°<br />

130°<br />

<strong>Gamma3</strong> Lag Screw<br />

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong><br />

Fig. 1


Features<br />

Fig. 5<br />

Lag Screw Stabilization System<br />

Lag Screw and<br />

Set Screw Function<br />

The Lag Screws are designed to<br />

transfer the load of the femoral head<br />

into the nail shaft by bridging the<br />

fracture line to allow fast and secure<br />

fracture healing. The load carrying<br />

thread design of the <strong>Gamma3</strong> Lag<br />

Screw provides large surface contact to<br />

the cancellous bone. This provides<br />

high resistance against cut out.<br />

<strong>Gamma3</strong> Lag Screws feature a special<br />

tip profile to allow use with bone<br />

substitutes and the self-tapping thread<br />

is designed for easy insertion.<br />

The patented Set Screw is designed to<br />

fit into one of the four grooves of the<br />

shaft of the Lag Screw. This prevents<br />

both, rotation and medial migration<br />

of the Lag Screw.<br />

The nail allows sliding of the Lag<br />

Screw to the lateral side for dynamic<br />

bone compression at the fracture site<br />

to enhance fracture healing.<br />

Technical Specifications<br />

• Lag Screw diameter: 10.5mm<br />

• Lag Screw lengths: 70−120mm in<br />

5mm increments<br />

• Patented Lag Screw design for high<br />

load absorption and easy insertion<br />

• Asymmetrical depth profile to allow<br />

the Lag Screw to slide in the lateral<br />

direction only (see orange arrow on<br />

Fig. 5).<br />

• Patented self retaining Set Screw to<br />

secure the Lag Screw against<br />

rotation and simultaneously<br />

allowing sliding of the Lag Screw<br />

laterally.


Features<br />

Distal Locking Screws<br />

The distal Locking Screw has a short<br />

self-tapping tip which facilitates a<br />

faster and easier start as well as easy<br />

screw insertion. It promotes excellent<br />

surface to bone contact (Fig. 6).<br />

Fig. 6<br />

The screw has an external diameter<br />

of 5mm, and provides an even higher<br />

fatigue strength than the clinically<br />

successful 6.28mm Locking Screw<br />

of the regular Gamma and G /K<br />

Locking <strong>Nail</strong> System (data on file).<br />

The screw diameter directly under the<br />

screw head has been reduced to<br />

prevent radial pressure that may cause<br />

micro fractures during screw insertion<br />

when the screw head reaches its final<br />

position. This reduction in diameter<br />

also improves the feel on the final<br />

tightening of the screw (Fig. 6a).<br />

Length Definition of<br />

the Distal Locking Screw<br />

The distal Locking Screw is measured<br />

from head to tip (Fig. 6b).<br />

Technical Specifications<br />

• Distal Locking Screw Diameter:<br />

5mm.<br />

• Distal Locking Screw lengths<br />

ranging from 25−50mm, in 2.5 and<br />

5mm increments. <strong>Long</strong>er screws up<br />

to 120mm are available on request.<br />

• Fully threaded screw design.<br />

Partially threaded screws are<br />

available on request<br />

• Self-tapping screw tip with<br />

optimized short cutting flutes.<br />

• Optimized diameter under the<br />

head helps to prevent microfractures<br />

during insertion.<br />

5mm<br />

Length Definition<br />

Reduced diameter<br />

Fig. 6a<br />

Fig. 6b


Features<br />

<strong>Gamma3</strong> System Benefits<br />

Strength and Stability<br />

The biomechanical superiority of the<br />

intramedullary system offers significantly<br />

greater strength and stability<br />

compared with the side plate, in<br />

clinical use [1]. The new <strong>Gamma3</strong><br />

system offers the same strength as the<br />

well established Gamma Locking <strong>Nail</strong><br />

System.<br />

[1] K.S.Leung et al, Multicenter Trial<br />

of the Modified Gamma <strong>Nail</strong> in East<br />

Asia CORR323: 146−154,1996<br />

D<br />

D > d<br />

d<br />

The Biomechanical Advantage<br />

over Side-Plate Systems<br />

Since the load-bearing axis of the<br />

<strong>Gamma3</strong> <strong>Nail</strong> is closer to the hip joint<br />

fulcrum, the effective lever arm on the<br />

implant and femur is significantly<br />

shorter than with an extramedullary<br />

plate. The reduction factor is equivalent<br />

to d/D as shown in Figure 7 (approximately<br />

25% [1]).<br />

The resultant force is transmitted<br />

directly down the femur using a nail<br />

system. If a side-plate system is used,<br />

the femur shaft may be weakened<br />

through a high amount of locking<br />

screws. The <strong>Gamma3</strong> <strong>Nail</strong> increases<br />

both the strength and reliability of the<br />

bio-mechanical repair. The distal<br />

dynamic locking option additionally<br />

allows the use of dynamic<br />

compression.<br />

Fig. 7<br />

Rehabilitation Benefits<br />

The extra strength effectively gained<br />

through the biomechanics of the<br />

<strong>Gamma3</strong> System combined with<br />

improved control of axial telescoping<br />

and rotational instability may allow<br />

earlier weight-bearing even in patients<br />

with complex or unstable proximal<br />

and combined ipsilateral shaft<br />

fractures. Early mobilization, dynamic<br />

compression, and a less traumatic<br />

operative technique increase the<br />

chance for rapid recovery and reliable<br />

bone union.


<strong>Operative</strong> <strong>Technique</strong><br />

Indications /Contraindications<br />

Indications<br />

• Subtrochanteric fractures<br />

• Pertrochanteric fractures associated<br />

with shaft fractures<br />

• Pathological fractures (including<br />

prophylactic use) in both trochan-<br />

teric and diaphysal areas<br />

• Nonunion and malunion<br />

Contraindications<br />

Contraindications are medial neck<br />

fractures.<br />

Note:<br />

If no bone consolidation occurs the<br />

system may fail. The aim of postoperative<br />

care must be to ensure the<br />

promotion of bone consolidation.<br />

The aim of this operative technique<br />

manual is to provide the surgeon with<br />

a simple step-by-step operating guide<br />

to aid in successful addition of the<br />

<strong>Gamma3</strong> System into their standard<br />

trauma care. Once the technique has<br />

been learned, the surgeon should find<br />

the operative procedure simple to<br />

implement. In fact, many of the basic<br />

principles for the <strong>Gamma3</strong> System<br />

are those employed for all closed intramedullary<br />

nailing procedures.<br />

This operative technique has been<br />

devised in consultation with leading<br />

surgeons in many countries to be<br />

a basic guide, particularly for less<br />

experienced users of the <strong>Gamma3</strong><br />

System. It is acknowledged that<br />

several alternative approaches to certain<br />

elements of the procedure are<br />

available, and may have advantages for<br />

particular situations or surgeons.<br />

Fig. 8


<strong>Operative</strong> <strong>Technique</strong><br />

Fig. 9<br />

X-ray in a-p view,<br />

showing implant<br />

Implant Selection<br />

The <strong>Gamma3</strong> <strong>Nail</strong> with a 125° nail<br />

angle may be used in the majority of<br />

patients. The 120° nail may be needed<br />

in patients with osteoarthritic coxa<br />

vara, and the 130° nail for coxa valga.<br />

Where such variations in femoral<br />

anatomy require an alternative, the<br />

following chapter describes how to<br />

select the optimal implant size.<br />

Preoperative Planning<br />

X-ray templates are very helpful<br />

during preoperative planning. Use the<br />

X-ray Templates (Fig. 9a) for short and<br />

long nails to select the correct implant<br />

and the optimal nail angle.<br />

These templates show the true implant<br />

size at a magnification of 15% in<br />

anterior-posterior view. The X-rays<br />

should be taken at this magnification<br />

(15%) for an optimum surgical<br />

outcome (see Fig. 9). If accurate anatomical<br />

reduction has been achieved,<br />

the X-ray can be taken from the fractured<br />

hip or from the contralateral<br />

side.<br />

Alternatively the femoral neck angle,<br />

i.e. the angle between the femoral<br />

shaft mid-axis and the femoral neck<br />

mid-axis, could be measured using a<br />

goniometer. The nail length may also<br />

be determinated intraoperatively using<br />

the Guide Wire Ruler together with<br />

the Guide Wire.<br />

10<br />

Fig. 9a<br />

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong> X-ray Template<br />

(Ref. No 1320-0005)<br />

Note:<br />

Please ensure precise alignment of<br />

the affected hip joint when using<br />

these templates. Template magnification<br />

is 15 %. All dimensions (nail<br />

angle and implant sizing) resulting<br />

from using these templates must be<br />

verified intraoperatively to ensure<br />

proper implant selection.


<strong>Operative</strong> <strong>Technique</strong><br />

Patient Positioning and<br />

Fracture Reduction<br />

The procedure for patient positioning<br />

is usually similar to that of a subtrochanteric<br />

fracture, however, in fractures<br />

that are particularly difficult to<br />

reduce, a transcondylar sterile Steinmann<br />

pin may be used. The pin is<br />

fixed directly to the orthopaedic table<br />

by an adaptable stirrup, and traction is<br />

applied until anatomical reduction in<br />

the A-P view is obtained (Fig. 10).<br />

The patient is placed in a supine position<br />

on the fracture table and closed<br />

reduction of the fracture is recommended<br />

(Fig. 10a).<br />

Traction is applied to the fracture,<br />

keeping the leg straight. The upper<br />

trunk should be flexed to the opposite<br />

side so that the fracture can be<br />

reduced by not too much adduction of<br />

the lower limb. It also gives easy access<br />

to the greater trochanter.<br />

11<br />

Fig. 10<br />

Fig. 10a


<strong>Operative</strong> <strong>Technique</strong><br />

1<br />

Fig. 11<br />

Fig. 11a<br />

Patient Positioning and<br />

Fracture Reduction<br />

Maintaining traction, the leg is internally<br />

rotated 10−15 degrees to complete<br />

fracture reduction; the patella<br />

should have either a horizontal or<br />

slightly inward position (Fig. 11).<br />

Position the image intensifier so that<br />

anterior-posterior and mediolateral<br />

views of the trochanteric region of the<br />

affected femur can be easily obtained.<br />

This position is best achieved if the<br />

image intensifier is positioned so that<br />

the axis of rotation of the intensifier is<br />

centered on the femoral neck of the affected<br />

femur (Fig. 11a). The views of<br />

the distal femur should also be easily<br />

obtained for distal locking.<br />

It is important to ensure that a view<br />

of both the distal and proximal tips<br />

of the nail can be obtained during the<br />

procedure without obstruction by the<br />

traction table.<br />

The patient is then prepared and<br />

draped as for standard femoral nailing<br />

procedures. When positioning the<br />

drapes, bear in mind that the incision<br />

will be more proximal.<br />

Note:<br />

Reduction should be achieved as anatomically as possible. If this is not<br />

achievable, reduction should be achieved at least in one plane. Reduction in the<br />

other plane may be achieved with the <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong> during insertion.


<strong>Operative</strong> <strong>Technique</strong><br />

Special <strong>Technique</strong>s for<br />

Fracture Reduction<br />

For specific situations, special<br />

techniques have been developed<br />

for fracture reduction, that are<br />

explained below.<br />

To counter this misalignment, the<br />

trunk is turned to the opposite side<br />

and held in position by a thoracic rest<br />

or by a large drape. This tightens the<br />

gluteus medius muscles and relaxes<br />

the psoas, externally rotating the<br />

proximal fragment into alignment<br />

and exposing the trochanter for easier<br />

introduction of the nail. The fractured<br />

limb is kept straight, with the knee in<br />

flexion (Fig. 12), using the stirrup to<br />

avoid adduction. This position helps<br />

to align the distal portion. Reduction<br />

is confirmed in the AP view.<br />

Subtrochanteric fractures cannot<br />

always be reduced during positioning<br />

in the lateral view, because the<br />

proximal fragment is drawn forward<br />

by tension from the psoas muscles.<br />

This may be reduced during surgery<br />

by using the Universal Rod (Fig. 12a).<br />

Care must be taken when introducing<br />

the implant as the proximal fragment<br />

may rotate during insertion.<br />

13<br />

Fig. 12<br />

Fig. 12a


<strong>Operative</strong> <strong>Technique</strong><br />

1<br />

Fig. 13<br />

Fig. 14<br />

Fig. 15<br />

Fig. 16<br />

Incision<br />

Incisions may be developed in<br />

different manners. Two alternatives<br />

will be described below.<br />

Alternative 1:<br />

The tip of the greater trochanter may<br />

be located by palpation (Fig. 13) and<br />

a horizontal skin incision of approximately<br />

2−3cm is made from the<br />

greater trochanter in the direction<br />

of the iliac crest (Fig. 14). In obese<br />

patients the incision length may need<br />

to be longer, depending on obesity of<br />

the patient.<br />

A small incision is deepened through<br />

the fascia lata, splitting the abductor<br />

muscle approximately 1−2cm immediately<br />

above the tip of the greater trochanter,<br />

thus exposing its tip. A selfretaining<br />

retractor, or tissue protection<br />

sleeve is put in place.<br />

Alternative :<br />

A long and thin metal rod (e. g. Screw<br />

Scale, <strong>Long</strong>) is placed on the lateral side<br />

of the leg. Check with image intensifier,<br />

using l-m view, that the metal rod<br />

is positioned parallel to the bone in the<br />

center of the proximal part of the<br />

femoral canal (Fig. 16a). A line is drawn<br />

on the skin (Fig. 16).<br />

Fig. 16a


<strong>Operative</strong> <strong>Technique</strong><br />

The C-arm is turned approx. 90° to<br />

provide an A-P image of the tip of<br />

the trochanter using the metal rod as<br />

shown in Fig. 17 and 17a.<br />

Fig. 17a<br />

A vertical line is drawn onto the skin<br />

(Fig. 18). The intersection of the lines<br />

indicates the position for the entry<br />

point of the nail. This is usually the<br />

anterior third of the tip of the greater<br />

trochanter as shown in Fig. 22.<br />

The skin incision is made cranially to<br />

the indicated intersection, following<br />

the sagital line in cranial direction.<br />

The distance between the intersection<br />

and the starting point for the incision<br />

differs, depending on the obesity of<br />

the patient. Under normal conditions<br />

it is a distance of approximately 2 cm.<br />

A small skin incision is made as<br />

described in Alternative 1 and shown<br />

in Fig. 20.<br />

1<br />

Fig. 17<br />

Fig. 18<br />

Fig. 19<br />

Fig. 20


<strong>Operative</strong> <strong>Technique</strong><br />

anterior 1 /3 2 /3 posterior<br />

1<br />

Fig. 21<br />

Fig. 22<br />

Fig. 23<br />

Incision<br />

Using a finger, the tip of the trochanter<br />

should be felt easily (Fig. 21).<br />

Entry Point<br />

The correct entry point is located at<br />

the junction of the anterior third and<br />

posterior two-thirds of the tip of the<br />

greater trochanter and on the tip itself<br />

(Fig. 22).<br />

Preparation of the<br />

Medullary Canal<br />

In order to prepare the medullary<br />

canal for the <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong>,<br />

3 possibilities are described in the next<br />

chapters.<br />

Alternative 1:<br />

Opening the Cortex<br />

The medullary canal has to be opened<br />

under image intensification. The use<br />

of the cannulated Curved Awl (Fig. 23)<br />

is recommended if conventional<br />

reaming or the One Step Conical<br />

Reamer will be used to prepare the<br />

canal for the nail.


<strong>Operative</strong> <strong>Technique</strong><br />

Reaming the Medullary Canal<br />

A 3mm ball-tipped Guide-Wire is<br />

recommended as a reamer guide.<br />

Pass the reamer Guide Wire through<br />

the cannulated curved awl into the<br />

shaft of the femur as shown, using the<br />

Guide Wire Handle (Fig. 24).<br />

Rotating the Guide Wire during insertion<br />

makes it easier to achieve the<br />

desired position in the middle of the<br />

medullary canal.<br />

Flexible reamers are used to ream the<br />

shaft of the femur in stages starting<br />

from 9mm diameter and increasing<br />

in 0.5mm increments (Fig. 25). The<br />

canal should be reamed at least 2mm<br />

larger than the distal diameter of the<br />

nail, 13mm for the <strong>Gamma3</strong> <strong>Long</strong><br />

<strong>Nail</strong> (Fig. 26).<br />

In order to accommodate the proximal<br />

part of the <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong>,<br />

the subtrochanteric region must be<br />

opened up to 15.5mm (Fig. 27). This<br />

can be done either by reaming with<br />

the <strong>Stryker</strong> BIXCUT Reaming<br />

System (Fig. 25) or, alternatively, with<br />

the One Step Conical Reamer. For<br />

soft tissue protection, the Conical<br />

Reamer Sleeve should be used during<br />

reaming.<br />

Care must be taken with flexible<br />

reamers to ensure that the Guide-<br />

Wire is not displaced laterally during<br />

reaming. This could lead to resection<br />

of more bone on the lateral side,<br />

which in turn would lead to an offset<br />

position for the nail and a risk of<br />

shaft fracture.<br />

Note:<br />

Where the shaft is comminuted,<br />

reaming should be stopped at<br />

the fracture site and penetration<br />

continued with the power drill off.<br />

Bixcut Reamer<br />

The complete range of Bixcut<br />

reamers is available with either<br />

modular or fixed heads.<br />

13mm<br />

13mm 15,5mm<br />

1<br />

approx. 80mm<br />

Fig. 25<br />

Fig. 24<br />

Fig. 26<br />

Fig. 27


<strong>Operative</strong> <strong>Technique</strong><br />

3mm Guide Wire<br />

or 3.2mm K-Wire One Step Conical Reamer<br />

K-Wire<br />

Multi Hole Trocar<br />

1<br />

Reamer Sleeve<br />

Fig. 28<br />

Fig. 29<br />

Alternative :<br />

One Step Conical Reamer<br />

The One Step Conical Reamer is an<br />

optional instrument and has been<br />

developed to provide surgeons with<br />

another option to prepare the<br />

proximal canal of the trochanter<br />

using only one drilling step.<br />

Using <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong>s, it is<br />

recommended to ream the medullary<br />

canal all the way down to the condyle<br />

area, at least up to a diameter of<br />

13mm.<br />

After skin incision and positioning<br />

of the Guide Wire as described<br />

above, the Trocar or Multi Hole<br />

Trocar is inserted into the Reamer<br />

Sleeve to protect the soft tissue<br />

during insertion. Push the Trocar<br />

(use center hole, if Multi Hole Trocar<br />

is used) and Sleeve Assembly down<br />

over the 3mm Guide Wire to the tip<br />

of the trochanter (Fig. 28 and 29).<br />

Entry Point Optimization<br />

The Entry Portal can also be made<br />

without using the awl. A 3.2mm<br />

K-Wire is placed through the tip of<br />

the trochanter.<br />

If you find that the K-Wire is not<br />

positioned in the optimal position, it<br />

may easily be corrected using a second<br />

K-Wire in combination with the<br />

Multi Hole Trocar.<br />

The Multi Hole Trocar has a special<br />

design for more precise insertion. In<br />

addition to the central hole, 4 other<br />

holes are located eccentrically at<br />

different distances from the center<br />

(Fig. 29) to easily revise insertion<br />

of the guiding K-Wire in the proper<br />

position (Entry Point).


<strong>Operative</strong> <strong>Technique</strong><br />

The Trocar is then removed and<br />

the One Step Conical Reamer is<br />

connected to the T-handle and slid<br />

over the Guide or K-Wire to the<br />

tip of the trochanter. With gentle<br />

clockwise turning and pushing<br />

movements, the Conical Reamer<br />

will drill into the proximal part of<br />

the trochanter (Fig. 30 and 31) and<br />

prepare the canal for the proximal<br />

part of the <strong>Gamma3</strong> <strong>Nail</strong>. The One<br />

Step Conical Reamer stops when the<br />

correct depth is reached.<br />

Note:<br />

The One Step Conical Reamer is a<br />

front and side cutting instrument<br />

and should be used with great care<br />

to ensure that the sharp edges of the<br />

reamer do not damage intact bone<br />

inadvertently.<br />

If a 3.2mm K-Wire was used it<br />

should be replaced by a Guide Wire<br />

afterwards.<br />

1<br />

Fig. 30<br />

Fig. 31


<strong>Operative</strong> <strong>Technique</strong><br />

0<br />

Fig. 32<br />

Alternative 3:<br />

Cannulated Cutter<br />

Opening the cortex<br />

The Cannulated Cutter is a front<br />

cutting device used to prepare the<br />

proximal part of the femur for the<br />

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong>.<br />

It provides surgeons with an advanced<br />

option to open the proximal femur<br />

cavity without reaming. Especially in<br />

older patients it may reduce the<br />

requirement for reaming of the<br />

femoral cavity. For the <strong>Long</strong> <strong>Nail</strong>, it is<br />

recommended to ream the complete<br />

femur all the way down to the condyle<br />

area up to a diameter of at least<br />

13mm.<br />

The Cannulated Cutter is guided over<br />

a solid 4mm Guide Pin. The fixation<br />

of this Guide Pin in the bone allows<br />

for an optimal placement for the<br />

Cannulated Cutter.<br />

This device allows for easy collection<br />

of bone graft material which might be<br />

helpful in difficult healing conditions.<br />

Fig. 33


<strong>Operative</strong> <strong>Technique</strong><br />

Assembly of Targeting Device<br />

1. Targeting Sleeve and<br />

Knob Assembly<br />

First assemble the Knob to the<br />

Targeting Sleeve (Fig. 34) and adjust<br />

the point on the Knob to be in line<br />

with the arrow on the Targeting<br />

Sleeve. Push the knob hard against<br />

the sleeve (Fig. 34a). The Knob moves<br />

approximately 5mm to the sleeve<br />

and has to be turned clockwise by<br />

approximately 30 degrees. Release the<br />

Knob and it will slip back the same<br />

distance. Now the Knob is assembled<br />

to the Targeting Sleeve and has to be<br />

connected to the Targeting Arm<br />

(Fig. 34b).<br />

. Targeting Arm and Targeting<br />

Sleeve Assembly<br />

Slide the Sleeve assembly over the<br />

Targeting Arm along the line until it<br />

stops (arrow line to arrow line).<br />

Rotate the Targeting Sleeve around to<br />

the required nail angle position for the<br />

Lag Screw, e. g. 125° (point to point) or<br />

distal locking positions, either<br />

“Dynamic” or “Static”. Now the<br />

Targeting Sleeve must be fixed in this<br />

position by pushing it strongly against<br />

the Targeting Arm. You will feel and<br />

hear, as the sleeve snaps into position.<br />

The Knobs only function is to lock the<br />

Lag Screw Guide Sleeve and the Tissue<br />

Protection Sleeve.<br />

Note:<br />

The Knob has to be assembled first<br />

to the Targeting Sleeve (Fig. 34a),<br />

otherwise the locking function of the<br />

sleeve may not work properly.<br />

Targeting Arm<br />

Targeting Sleeve 180<br />

green coded<br />

Knob<br />

Fig. 34<br />

1<br />

Fig. 34a Fig. 34b


<strong>Operative</strong> <strong>Technique</strong><br />

Ball Tip Screwdriver<br />

<strong>Nail</strong> Holding Screw<br />

Targeting Arm<br />

<strong>Gamma3</strong> <strong>Nail</strong> <strong>Long</strong><br />

Fig. 35<br />

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong> Assembly<br />

3. Assembly of the Targeting<br />

Device and the <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong><br />

The selected <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong> is<br />

now assembled to the Carbon Fibre<br />

Targeting Device as shown in Figure<br />

35. The nail connecting part of the<br />

Targeting Device is designed with an<br />

easy assembly function for fast and<br />

secure nail fixation.<br />

Ensure that the locating pegs fit into<br />

the corresponding notches of the<br />

proximal part of the nail.<br />

Fully tighten the <strong>Nail</strong> Holding Screw<br />

with the Ball Tip Screwdriver, so<br />

that it does not loosen during nail<br />

insertion.<br />

Before starting surgery the following<br />

two functions of the Targeting Device<br />

have to be checked:<br />

1. Secure fixation between <strong>Nail</strong> and<br />

Targeting Device<br />

2. Lag Screw Guide Sleeve matches<br />

the selected nail angle.


<strong>Operative</strong> <strong>Technique</strong><br />

Before checking the function of the<br />

Lag Screw Guide Sleeve, the Knob<br />

must be positioned in the counter<br />

clockwise position. Pass the Lag Screw<br />

Guide Sleeve gently through the hole<br />

of the Targeting Sleeve and tighten it<br />

gently in its final position by turning<br />

the Knob clockwise. Check correct nail<br />

angle using the K-Wire, 4.2mm Drill<br />

or Lag Screw Step Drill (Fig. 36).<br />

Remove the Lag Screw Guide Sleeve by<br />

turning the Knob counter clockwise<br />

and pulling the sleeve back.<br />

Note:<br />

Before starting surgery, the implant<br />

and instrument assembly has<br />

to be checked. Ensure that the<br />

Targeting Sleeve angle matches the<br />

corresponding nail angle chosen, e. g.<br />

a 125° Targeting Sleeve for a 125° nail<br />

(Fig 36).<br />

Lag Screw Guide Sleeve<br />

3<br />

Fig. 36


<strong>Operative</strong> <strong>Technique</strong><br />

Fig. 37<br />

Fig. 38<br />

Fig. 39<br />

Fig. 40<br />

Fig. 41<br />

<strong>Nail</strong> Insertion and Positioning<br />

Insert the <strong>Gamma3</strong> <strong>Nail</strong> by hand<br />

(Fig. 37).<br />

DO NOT use undue force −<br />

NEVER use a hammer for<br />

nail insertion.<br />

The final nail depth is monitored with<br />

the image intensifier; the axis of the<br />

Lag Screw may be projected with a<br />

ruler on the monitor screen to ensure<br />

that the Lag Screw is placed in the<br />

optimal position.<br />

Proceed until the axis of the Lag<br />

Screw hole (visible as a crescent shape<br />

on the screen) is aligned with the<br />

lower half of the femoral neck (Fig.<br />

38). The objective of this is to ultimately<br />

position the Lag Screw slightly<br />

inferior in the femoral head in the<br />

frontal plane.<br />

Note:<br />

Make sure to remove the Guide Wire<br />

using the Guide Wire Handle (Fig. 39).<br />

When the <strong>Gamma3</strong> <strong>Nail</strong> has been<br />

inserted to its final depth, check the<br />

anteversion of the nail. Use of the<br />

K-Wire Clip (Fig. 40) or the “One Shot<br />

Device” is recommended (see next<br />

page).<br />

Lag Screw Positioning<br />

using the K-Wire Clip<br />

The K-Wire Clip is mounted in the<br />

slots of the Targeting Arm by pressing<br />

the Clip flanges together.<br />

Using image intensification of the<br />

lateral view, the projection of the<br />

U-Wire on the bone shows the A-P<br />

positioning of the Lag Screw.<br />

The Lag Screw should be placed in<br />

the central position of the femoral<br />

head(Fig. 41).<br />

Before proceeding ensure that the <strong>Nail</strong><br />

Holding Screw is still fully tightened.


<strong>Operative</strong> <strong>Technique</strong><br />

Lag Screw Positioning using<br />

the One Shot Device<br />

The One Shot Device is<br />

recommended for optimal Lag<br />

Screw placement:<br />

The One Shot Device is recommended<br />

for verifying that the Lag<br />

Screw is placed in its optimal<br />

position. This device enables correct<br />

positioning of the K-Wire for Lag<br />

Screw placement before performing<br />

lateral skin incision and opening of<br />

the lateral cortex. Figures 42−43a<br />

give an overview of the working<br />

principle of the One Shot Device.<br />

nail positioned<br />

too caudal<br />

optimal nail position<br />

nail positioned<br />

too cranial<br />

Fig. 42a<br />

A/P view<br />

Positioning of nail depth<br />

Positioning of anteversion<br />

Fig. 42<br />

Fig. 43<br />

Fig. 43a<br />

Lateral view


<strong>Operative</strong> <strong>Technique</strong><br />

Fig. 44<br />

Fig. 45<br />

Lag Screw Guide Sleeve in good contact to the lateral cortex<br />

Lag Screw Insertion<br />

The Targeting Device may be held<br />

by an assistant to prevent it from<br />

externally rotating the nail until the<br />

next stage is completed.<br />

Next, assemble the Lag Screw Guide<br />

Sleeve and the green coded 4.2mm<br />

Lag Screw Drill Guide Sleeve and pass<br />

them through the Targeting Sleeve<br />

to the level of the skin. At the point<br />

of contact, perform a small incision<br />

down to the bone (Fig. 44). The Guide<br />

Sleeve assembly is then advanced<br />

through the incision. If the sleeves<br />

catches the fascia lata, twisting it will<br />

usually allow it to pass through to the<br />

bone.<br />

In order to assure accurate Lag Screw<br />

length measurement, the outer<br />

Guide Sleeve must be in good contact<br />

to the lateral cortex of the femur<br />

(Fig. 45). The Knob of the Target<br />

Sleeve must be turned gently clockwise<br />

to lock the Guide Sleeve in place<br />

and further stabilize the targeting<br />

assembly.


<strong>Operative</strong> <strong>Technique</strong><br />

With the Lag Screw Guide Sleeve<br />

firmly engaged in the cortex, the green<br />

coded 4.2mm Lag Screw Drill Guide<br />

Sleeve should be pushed gently against<br />

the cortex. Using the green coded<br />

4.2mm × 300mm center tipped drill,<br />

the lateral cortex should be opened by<br />

power tool or by hand (Fig. 46).<br />

The green coded 4.2mm Lag Screw<br />

Drill Guide Sleeve is then replaced by<br />

the K-Wire Sleeve.<br />

Both sleeves look similar but have<br />

different inner hole diameters. The K-<br />

Wire Sleeve has no colored ring.<br />

K-Wire Sleeve<br />

Lag Screw Drill Guide Sleeve<br />

Note:<br />

Before proceeding, check that the<br />

Guide Wire for the flexible reamer<br />

and nail insertion used earlier has<br />

been removed.<br />

The single use K-Wire inserted<br />

through the K-Wire Sleeve should<br />

be advanced up to the subchondral<br />

bone (Fig. 48), using the Guide Wire<br />

Handle. Check that the K-Wire is<br />

placed in the lower half of the femoral<br />

head in the frontal plane and on the<br />

midline in the lateral plane (Fig. 48).<br />

Check the position with the image<br />

intensifier in both the anteriorposterior<br />

and mediolateral views as<br />

shown in Figure 38 to ensure optimal<br />

K-Wire positioning.<br />

uSE K-WIRE FOR ONE<br />

SuRGICAL PROCEDuRE ONLy<br />

Fig. 46<br />

Opening of the lateral cortex<br />

Fig. 47<br />

K-Wire placement<br />


<strong>Operative</strong> <strong>Technique</strong><br />

Fig. 49<br />

Lag Screw length measurement<br />

Fig. 50<br />

Lag Screw Length Measurement<br />

Lag Screw Insertion<br />

The objective is to position the Lag<br />

Screw below the center of the femoral<br />

head in the anterior-posterior view<br />

and centrally in the lateral view, to<br />

provide the best load transfer to the<br />

Lag Screw.<br />

After satisfactorily positioning the<br />

K-Wire, the required Lag Screw length<br />

is measured using the Lag Screw Ruler.<br />

Before starting to measure, ensure<br />

that the Lag Screw Guide Sleeve is still<br />

pressed firmly against the lateral cortex<br />

of the femur (Fig. 49).<br />

Place the Lag Screw Ruler directly<br />

under the K-Wire (Fig. 50).<br />

The recommended value for the Step<br />

Drill depth and the Lag Screw length<br />

can be read directly off the Lag Screw<br />

Ruler. If the value is between markings<br />

on the scale, e. g. 97mm, it should<br />

always be rounded up to the next<br />

higher value, e. g. 100mm.<br />

Note:<br />

K-Wires are not intended for re-use.<br />

They are single use only. K-Wires may<br />

be damaged or bent during surgical<br />

procedures. If a K-Wire is re-used, it<br />

may get lodged in the drill and could<br />

be advanced into the pelvis during<br />

the following steps of the procedure.<br />

This may damage large blood vessels<br />

or cause other serious injuries.


<strong>Operative</strong> <strong>Technique</strong><br />

The value of the measurement (Fig.<br />

50) is now transferred to the adjustable<br />

stop on the Lag Screw Step Drill.<br />

The value e. g. 100 must be visible in<br />

the window of the Step Drill Stop (Fig.<br />

51).<br />

The K-Wire Sleeve is now removed<br />

and the adjusted Lag Screw Step Drill<br />

is passed over the K-Wire (Fig. 51a),<br />

through the Lag Screw Guide Sleeve.<br />

The channel for the Lag Screw is<br />

prepared using the T-handle<br />

connected to the Lag Screw Step Drill.<br />

If exceptional resistance is encountered,<br />

a power drill may be used with<br />

great care.<br />

Drilling should continue until the stop<br />

of the Step Drill comes into contact<br />

with the Lag Screw Guide Sleeve (Fig.<br />

51a). Ensure that the Targeting Device<br />

is well supported to prevent it from<br />

slipping back or rotating.<br />

The drilling process, especially<br />

when the tip of the drill comes close<br />

to its final position in the femoral<br />

head, should be controlled under an<br />

image intensifier to avoid hip joint<br />

penetration. The K-Wire can also be<br />

observed in the K-Wire window of the<br />

Step Drill.<br />

Note:<br />

It is important to observe the K-Wire<br />

tip during drilling on the intensifier.<br />

The K-Wire window provides an<br />

additional possibility to double check<br />

the K-Wire end position. Ensure that<br />

under no circumstances the K-Wire is<br />

advanced into the pelvis.<br />

K-Wire end<br />

Lock<br />

K-Wire window<br />

Fig. 51<br />

Window of the Step Drill Stop<br />

K-Wire window<br />

Fig. 51a<br />

Lag Screw<br />

Guide Sleeve<br />

Lag Screw Step<br />

Drill Stop<br />

Groove indicates<br />

K-Wire end position


<strong>Operative</strong> <strong>Technique</strong><br />

30<br />

Fig. 52<br />

Fig. 53<br />

Lag Screw and Lag Screwdriver assembly<br />

Lag Screw Insertion<br />

During drilling, monitor the depth of<br />

the drill near the subchondral bone<br />

on image intensification. At this stage,<br />

you should see the tip of the K-Wire<br />

protruding about 6 to 10mm out of the<br />

step drill (Fig. 52) because the threaded<br />

portion of the K-Wire was intentionally<br />

not included in the drill measurement.<br />

This is to prevent the drill from<br />

penetrating the joint and to ensure<br />

that the K-Wire remains anchored in<br />

the subchondral bone after reaming.<br />

Remove the Step Drill by turning it<br />

clockwise and pulling it backwards.<br />

The length of the Lag Screw chosen<br />

should be the same as that of the Step<br />

Drill (in this example 100mm). The<br />

screw is then assembled to the Lag<br />

Screwdriver (Fig. 53).<br />

In a case where compression is to<br />

be applied, a shorter Lag Screw<br />

length should be chosen to avoid<br />

the Lag Screw sticking out too far<br />

out of the lateral cortex (see chapter<br />

Compression / Apposition below).<br />

Ensure that the pegs of the Lag<br />

Screwdriver are in the notches of the<br />

Lag Screw. The thumbwheel of the<br />

handle must be turned clockwise and<br />

finally tightened using the Ball Tip<br />

Screwdriver.<br />

The Lag Screw assembly is now passed<br />

over the K-Wire, through the Lag<br />

Screw Guide Sleeve and threaded up to<br />

the end of the predrilled hole into the<br />

femur head. Check the end position of<br />

the Lag Screw on the image intensifier.<br />

A double check of the end position is<br />

also possible with the indicator ring on<br />

the Lag Screw Screwdriver. The final<br />

position is obtained when the indicator<br />

ring reaches the end of the Lag Screw<br />

Guide Sleeve.


<strong>Operative</strong> <strong>Technique</strong><br />

Lag Screw Fixation<br />

The handle of the Lag Screwdriver<br />

must be either parallel or perpendicular<br />

(90°) to the Targeting Arm (Fig.<br />

55 on next page) to ensure that the Set<br />

Screw is able to fit into one of the four<br />

grooves of the Lag Screw shaft.<br />

The Set Screw alignment indicator<br />

will help to find the right position of<br />

the handle.<br />

If the T-handle is not perpendicular or<br />

parallel to the Targeting Arm, turn it<br />

clockwise until it reaches this position.<br />

NEVER TURN THE LAG SCREW<br />

COUNTER CLOCKWISE.<br />

If the K-Wire is inadvertently removed,<br />

the screw may still be inserted without<br />

it, provided that the Guide Sleeve is still<br />

in contact with the cortex.<br />

Note:<br />

It is strongly recommended to place<br />

the Lag Screw at the end of predrilled<br />

hole in order to provide maximal<br />

resistance against cut out. Never turn<br />

the Lag Screw counter clockwise after<br />

the final position is reached because<br />

otherwise the Lag Screw may lose full<br />

bony surface contact to its tip.<br />

Compression / Apposition<br />

If compression or apposition of the<br />

fracture gap is required, this can be<br />

achieved by gently turning the thumbwheel<br />

of the Lag Screwdriver clockwise<br />

against the Guide Sleeve (Fig. 54).<br />

Before starting compression, make<br />

sure that the Lag Screw Guide Sleeve is<br />

unlocked to allow its free sliding. To<br />

unlock the Lag Screw Guide Sleeve, the<br />

Knob of the Target Device has to be<br />

turned counter clock-wise. In<br />

osteoporotic bone care must be taken<br />

to prevent Lag Screw pullout in the<br />

femoral head. The Lag Screw should be<br />

chosen shorter depending on the<br />

expected amount of compression.<br />

31<br />

Fig. 54<br />

Compression / Apposition turning the thumbwheel clockwise.


<strong>Operative</strong> <strong>Technique</strong><br />

Set Screw Alignment Indicator<br />

3<br />

Fig. 56<br />

Set Screw insertion<br />

Fig. 55<br />

T-handle end position<br />

Lag Screw Fixation<br />

Note:<br />

The Set Screw must be used.<br />

The use of the Set Screw is not an<br />

option.<br />

Assemble the Set Screw to the Set<br />

Screw Driver. Insert the Set Screw<br />

as shown in Figure 56 along the<br />

opening of the post of the Targeting<br />

Device and advance it through the<br />

<strong>Nail</strong> Holding Screw.<br />

Push the Set Screw Driver down<br />

until you are sure, that the Set Screw<br />

engages the corresponding thread<br />

in the nail. During pushing down<br />

the assembly, you may feel a slight<br />

resistance.<br />

Turn the Screwdriver handle<br />

clockwise under continoues pressure.<br />

You may notice a slight resistance<br />

when turning the Set Screw. This is<br />

because the Set Screw thread is<br />

equipped with the “Nylstop” system<br />

to prevent spontaneous loosening.<br />

This is not the final position for the<br />

Set Screw.<br />

Keep on turning the Set Screw until<br />

you feel contact in one of the grooves<br />

of the Lag Screw.


<strong>Operative</strong> <strong>Technique</strong><br />

To verify the engagement the Set<br />

Screw in a groove of the Lag Screw,<br />

try to turn the Lag Screwdriver gently<br />

clockwise and counter-clockwise.<br />

If it is not possible to turn the Lag<br />

Screwdriver the Set Screw is engaged<br />

in one of the grooves. If the Lag Screw<br />

Driver still moves, recorrect the<br />

T-handle position and tighten the Set<br />

Screw again until it engages in one of<br />

the four Lag Screw grooves.<br />

After slightly tightening the Set Screw,<br />

it should then be unscrewed by one<br />

quarter (¼) of a turn, until a small<br />

play can be felt at the Lag Screwdriver.<br />

This ensures a free sliding of the Lag<br />

Screw.<br />

Make sure that the Set Screw is still<br />

engaged in the groove by checking that<br />

it is still not possible to turn the Lag<br />

Screw with the Lag Screwdriver.<br />

Note:<br />

Do not unscrew the Set Screw more<br />

than ¼ of a turn.<br />

If distal locking is not indicated, the<br />

End Cap should be assembled to the<br />

nail end to prevent bone ingrowth.<br />

Leaving the Lag Screwdriver in<br />

place, the <strong>Nail</strong> Holding Screw is<br />

now removed using the Ball Tip<br />

Screwdriver, Universal Socket Wrench<br />

or Spreading Screwdriver and turning<br />

it counter clockwise. Insert the End<br />

Cap (size 0) using the Socket Wrench,<br />

Spreading Screwdriver or the Ball Tip<br />

Screwdriver. The End Cap should be<br />

tightened slightly.<br />

Please see “End Cap Insertion” chapter.<br />

Alternatively the End Cap could also<br />

be inserted free hand after removal of<br />

the Target Device.<br />

33<br />

Fig. 57


<strong>Operative</strong> <strong>Technique</strong><br />

Fig. 58a<br />

Dynamic Locking<br />

Fig. 58b<br />

Secondary Dynamization<br />

Not in line with the <strong>Nail</strong> holes<br />

This is the best position to drill<br />

Not in line with the <strong>Nail</strong> holes<br />

This is the best position to drill; it shows correct view<br />

to be in line with the <strong>Nail</strong> holes<br />

3<br />

Fig. 58c<br />

Static Locking<br />

Fig. 58<br />

Fig. 59<br />

Fig. 60<br />

Distal Screw Locking<br />

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong>s offer the<br />

possibility to be locked distally. For<br />

distal locking, the <strong>Long</strong> <strong>Nail</strong> offers the<br />

following three locking options (Fig.<br />

58), depending on the fracture pattern.<br />

<strong>Long</strong> <strong>Nail</strong> Distal Locking Options<br />

• Locking in the distal part of the<br />

oblong hole creates a dynamic<br />

locking mechanism − requires only<br />

one screw (see Fig. 58a).<br />

• One screw placed in the distal part<br />

of the oblong hole and the other in<br />

the round hole.<br />

If dynamization is required after a<br />

period of time, the screw, placed in<br />

the round hole, has to be removed −<br />

requires two screws (see Fig. 58b).<br />

• One screw placed in the round<br />

hole and the other is placed in the<br />

proximal part of the oblong hole −<br />

requires two screws (see Fig. 58c).<br />

Distal locking is recommended:<br />

• if the fracture is unstable<br />

• if rotational stability is required<br />

• if there is a wide disparity between<br />

the diameter of the nail and the<br />

femoral cavity.<br />

Various techniques can be used to<br />

guide drilling and insertion of screws<br />

through the distal holes. The freehand<br />

technique is described below.<br />

Visualizing the distal holes<br />

The essential initial step in distal<br />

targeting is to position the image<br />

intensifier so that the distal hole in<br />

the nail appears perfectly round.<br />

Naturally, this visualization steps<br />

refer to the appearance of the round<br />

and not the oblong hole. If the hole<br />

appears to be elliptical in either the<br />

vertical or horizontal plane, the image<br />

intensifier position must be adjusted<br />

appropriately as shown in Figures 59<br />

and 60.<br />

It is advised to correct image in one<br />

plane at a time.


<strong>Operative</strong> <strong>Technique</strong><br />

Free-hand <strong>Technique</strong><br />

The free-hand drill technique is used<br />

to fix the distal bone fragment to the<br />

nail using Locking Screws. Length and<br />

rotational alignment of the leg must<br />

be checked before locking the nail.<br />

The distal nail locking is described<br />

as follows, using the Static Locking<br />

mode according to Figures 61−63. Skin<br />

incisions are made in line with the<br />

distal holes of the nail.<br />

Once the image intensifier is correctly<br />

positioned as shown in Figures<br />

59 and 60, use the centre tipped<br />

Ø4.2mm×180mm, green coded drill<br />

and place the tip of the drill at an<br />

oblique angle to the centre of the hole<br />

(Fig. 61). Verify the position by X-ray<br />

and move the drill into the same plane<br />

as the holes in the nail, then drill<br />

through the first cortex and the nail<br />

until resistance of the second cortex is<br />

felt as shown in Figure 62.<br />

Alternatively, the drill can be drilled<br />

through the second cortex while<br />

viewing the image intensifier. The<br />

screw length can then be read directly<br />

from the Screw Scale on the drill<br />

(Fig. 64).<br />

If the Tissue Protection Sleeve is used<br />

with the drill, it has to be removed for<br />

the measurement.<br />

It is also possible to measure the<br />

correct screw length using the Free<br />

Hand Screw Gauge. After drilling<br />

through the second cortex, remove the<br />

drill and advance the small hook of the<br />

Screw Gauge through the holes behind<br />

the medial cortex and read out the<br />

required locking screw length.<br />

Insert the 5mm distal Locking Screw<br />

through the skin by using the 3.5mm<br />

Screwdriver; advance the screw head<br />

carefully until it is just in direct<br />

contact with the cortex (Fig. 65).<br />

Note:<br />

Take care not to overtighten.<br />

The screw head should just come<br />

into contact with the cortex and<br />

resistance should be felt.<br />

Fig. 62<br />

add thickness of the cortex<br />

(approx +5mm) to the read out value<br />

3<br />

anterior<br />

posterior<br />

Fig. 61<br />

Fig. 63<br />

direct read out<br />

Fig. 64<br />

Fig. 65


<strong>Operative</strong> <strong>Technique</strong><br />

3<br />

Fig. 66<br />

Fig. 67<br />

Alternative<br />

Alternatively Condyle Screws could<br />

be used for distal locking. If a<br />

Condyle Screw will be inserted, both<br />

cortices are drilled to a diameter of<br />

5mm using the Ø5×230mm Drill in<br />

a free-hand drill technique.<br />

After drilling through the second<br />

cortex, remove the drill and advance<br />

the small hook of the Screw Gauge<br />

through the holes behind the medial<br />

cortex and read out the required<br />

condyle screw length (see Fig. 66).<br />

Note:<br />

The measurement equals Condyle<br />

Screw fixation length (from top of<br />

the Condyle Screw head to the top<br />

of Condyle Nut head, as shown in<br />

Fig. 66). The Condyle Screw length<br />

is defined with the Condyle Screw<br />

tip flush to the Condyle Nut head.<br />

The possible fixation length can<br />

be 2mm longer than the Condyle<br />

Screw length or 5mm shorter.<br />

Please ensure that the Condyle Nut<br />

is tightened a minimum of 5 turns<br />

on the Condyle Screw!<br />

Examples:<br />

Measurement Condyle Screw Length<br />

(mm) (mm)<br />

56 55<br />

57 55 or 60<br />

58 60<br />

59 60<br />

The Condyle Screw K-Wire<br />

Ø1.8×310mm inserted from the<br />

lateral side to the medial side. At the<br />

medial point of the perforation, a<br />

skin incision is made for the Condyle<br />

Screw. From the medial side, the<br />

Condyle Screw is now brought<br />

forward over the Condyle Screw K-<br />

Wire and inserted using the Condyle<br />

Screw Screwdriver.<br />

Insert the Condyle Nut over the K-<br />

Wire using the other Condyle Screw<br />

Screwdriver (Fig. 67).<br />

Alternatively, if patient anatomy<br />

allows, the Condyle Screw may be<br />

introduced from lateral to medial in a<br />

similar manner as described above.


<strong>Operative</strong> <strong>Technique</strong><br />

Using both Condyle Screw<br />

Screwdrivers, the Condyle Nut and the<br />

Condyle Screw are tightened. Once<br />

tightened, the K-Wire is removed.<br />

The adjustable screw washers of the<br />

Condyle Screw and the Condyle<br />

Nut adapt to the surface of the bone.<br />

3<br />

Fig. 68


<strong>Operative</strong> <strong>Technique</strong><br />

3<br />

Fig. 69<br />

End Cap assembly<br />

Fig. 70<br />

Final <strong>Nail</strong> assembly<br />

End Cap Insertion<br />

It is recommended to use an End Cap<br />

to close the proximal part of the nail to<br />

prevent bone ingrowth.<br />

Remove the <strong>Nail</strong> Holding Screw using<br />

the Ball Tip Screwdriver, Spreading<br />

Screwdriver, Universal Socket Wrench<br />

or Strike Plate. Load the End Cap (size<br />

0) to one of the Screwdrivers and pass<br />

the assembly through the top of the<br />

Targeting Device down into the nail.<br />

Turn the handle clockwise until<br />

it stops mechanically. Remove the<br />

Screwdriver and remove the Targeting<br />

Device in cranial direction.<br />

Alternatively the End Cap could also<br />

be inserted free hand after removal of<br />

the Targeting Device.


<strong>Operative</strong> <strong>Technique</strong><br />

<strong>Nail</strong> Extension End Caps<br />

If the proximal end of the nail is<br />

completely sunk in the trochanter,<br />

End Caps in size +5mm and +10mm<br />

are available and can be assembled to<br />

the nail instead of the End Cap size<br />

0, to achieve cortical bone support<br />

proximally. The proximal part of<br />

the nail will be elongated by 5mm or<br />

10mm.<br />

These elongation End Caps are<br />

assembled using the Strike Plate with<br />

the self-retaining ring, the Spreading<br />

Screwdriver or Ball Tip Screwdriver.<br />

This can only be done if the Targeting<br />

Device is already removed from the<br />

nail.<br />

Postoperative Care<br />

and Rehabilitation<br />

Active and passive mobilization of the<br />

lower limbs may be started<br />

immediately. The injured limb should<br />

be kept elevated.<br />

For stable fractures that are locked<br />

statically or dynamically, full weight<br />

bearing walking may be started<br />

immediately.<br />

For unstable fractures with static<br />

locking, immediate full weight bearing<br />

walking is allowed in fractures with<br />

good bone contact.<br />

For fractures with poor bone contact<br />

due to comminution, partial weightbearing<br />

walking is allowed for the<br />

first 6 to 8 weeks. Full weight bearing<br />

walking can be commenced when<br />

there is a bridging callus formed as<br />

evident on the follow up X-ray.<br />

End Cap (size +10mm)<br />

End Cap (size +5mm)<br />

3<br />

Fig. 71


<strong>Operative</strong> <strong>Technique</strong><br />

0<br />

Fig. 72<br />

Fig. 73<br />

Fig. 74<br />

Extraction of the<br />

<strong>Gamma3</strong> Implant<br />

Where implant extraction is indicated,<br />

please proceed as follows:<br />

Step I (Fig. )<br />

Remove the distal screw using the<br />

3.5mm Screwdriver after making an<br />

incision through the old scar.<br />

Step II (Fig. 3)<br />

Make a small incision through the<br />

old scar below the greater trochanter<br />

to expose the outer end of the Lag<br />

Screw. Remove any bony ingrowth<br />

which may be obstructing the outer<br />

end or internal thread of the Lag<br />

Screw as necessary to enable the Lag<br />

Screwdriver to engage fully.<br />

The K-Wire is then introduced via the<br />

Lag Screw into the head of the femur.<br />

The Lag Screwdriver is passed over the<br />

K-Wire, using the Lag Screw Guide<br />

Sleeve as a Tissue Protector, and<br />

engaged with the distal end of the<br />

Lag Screw.<br />

Check that ingrowth does not<br />

obstruct secure engagement of the<br />

Lag Screwdriver, otherwise the<br />

Lag Screw or Screwdriver may be<br />

damaged and extraction will be<br />

much more difficult. Tighten the<br />

thumbwheel clockwise.<br />

Step III (Fig. )<br />

An incision is made over the proximal<br />

end of the nail, the proximal End Cap<br />

if used is removed using the Ball Tip<br />

Screwdriver, Spreading Screwdriver or<br />

Strike Plate, and the Set Screwdriver<br />

is engaged with the Set Screw. The Set<br />

Screw is rotated anti-clockwise until it<br />

is removed.<br />

Note:<br />

As the targeting device is not<br />

connected to the nail, we recommend<br />

using the Straight Set Screwdriver<br />

(1320-0210) for better guidance<br />

through the soft tissue to get access to<br />

the Set Screw.


<strong>Operative</strong> <strong>Technique</strong><br />

Step IV (Fig. )<br />

The Conical Extraction Rod is then<br />

threaded and tightened into the<br />

proximal end of the nail. The Lag<br />

Screw is extracted by anti clockwise<br />

rotation and pulling of the Lag Screwdriver.<br />

The K-Wire must then be<br />

removed.<br />

Note:<br />

It is useful to first turn the Lag Screw<br />

Screwdriver clockwise slightly to<br />

loosen possibly bony ingrowth in<br />

the screw threads before turning it<br />

counter clockwise.<br />

Step V (Fig. & )<br />

An appropriate sliding hammer<br />

assembly is attached to the Extraction<br />

Rod and the nail extracted.<br />

1<br />

Fig. 75<br />

Fig. 76<br />

Fig. 77


<strong>Operative</strong> <strong>Technique</strong><br />

Fig. 78<br />

Dealing with Special Cases<br />

Posterior Displacement<br />

In case of a comminuted fracture,<br />

there is a tendency of the fracture to<br />

become displaced posteriorly, making<br />

it difficult to place the K-Wire into the<br />

center of the neck and head. This can<br />

be solved by lifting the nail insertion<br />

Targeting Device (Fig. 78).<br />

Alternatively, an assistant can lift up<br />

the greater trochanter manually or<br />

with a reduction spoon; or support it<br />

with a sandbag. This will maintain the<br />

neck and the femur in almost the same<br />

axis, facilitating passage of the K-Wire<br />

through the center of the neck and<br />

head.<br />

The position should then be checked in<br />

both the anterior-posterior and lateral<br />

views using the image intensifier.


Ordering Information – Implants<br />

Packaging<br />

All implants are packed sterile only.<br />

The <strong>Nail</strong> and Lag Screw Implant have<br />

to be secured using the Set Screw<br />

in every surgical operation, without<br />

exception (see also page 33).<br />

The <strong>Nail</strong> and the Set Screw are<br />

therefore supplied together in the<br />

same blister pack (see Fig. 79).<br />

The blister is packed in a white<br />

carton and wrapped to protect the<br />

contents during transportation and<br />

storage.<br />

Only two package sizes are used for<br />

all the nails (Fig. 80).<br />

The long nails are packed in a longer<br />

box and the short nails in a shorter<br />

box.<br />

This facilitates identification in the<br />

storage area.<br />

The package carries also the date<br />

of sealing and a sterility expiration<br />

date.<br />

<strong>Long</strong> <strong>Nail</strong>, packaging example<br />

Trochanteric <strong>Nail</strong>, packaging example<br />

<strong>Gamma3</strong> Set Screw<br />

<strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong><br />

Fig. 80 Fig. 79<br />

3


Ordering Information – Implants<br />

<strong>Long</strong> <strong>Nail</strong> Kit R .0, Ti, Left*, Ø1 . /11mm<br />

Lag Screw, TI**<br />

Titanium Length Angle<br />

REF mm °<br />

3320-0280S 280 120<br />

3320-0300S 300 120<br />

3320-0320S 320 120<br />

3320-0340S 340 120<br />

3320-0360S 360 120<br />

3320-0380S 380 120<br />

3320-0400S 400 120<br />

3320-0420S 420 120<br />

3320-0440S 440 120<br />

3320-0460S 460 120<br />

3325-0280S 280 125<br />

3325-0300S 300 125<br />

3325-0320S 320 125<br />

3325-0340S 340 125<br />

3325-0360S 360 125<br />

3325-0380S 380 125<br />

3325-0400S 400 125<br />

3325-0420S 420 125<br />

3325-0440S 440 125<br />

3325-0460S 460 125<br />

3330-0280S 280 130<br />

3330-0300S 300 130<br />

3330-0320S 320 130<br />

3330-0340S 340 130<br />

3330-0360S 360 130<br />

3330-0380S 380 130<br />

3330-0400S 400 130<br />

3330-0420S 420 130<br />

3330-0440S 440 130<br />

3330-0460S 460 130<br />

Titanium Diameter Length<br />

REF mm mm<br />

3060-0070S 10.5 70<br />

3060-0075S 10.5 75<br />

3060-0080S 10.5 80<br />

3060-0085S 10.5 85<br />

3060-0090S 10.5 90<br />

3060-0095S 10.5 95<br />

3060-0105S 10.5 100<br />

3060-0105S 10.5 105<br />

3060-0115S 10.5 110<br />

3060-0115S 10.5 115<br />

3060-0120S 10. 5 120<br />

<strong>Long</strong> <strong>Nail</strong> Kit R .0, Ti, Right*, Ø1 . /11mm<br />

Titanium Length Angle<br />

REF mm °<br />

3220-0280S 280 120<br />

3220-0300S 300 120<br />

3220-0320S 320 120<br />

3220-0340S 340 120<br />

3220-0360S 360 120<br />

3220-0380S 380 120<br />

3220-0400S 400 120<br />

3220-0420S 420 120<br />

3220-0440S 440 120<br />

3220-0460S 460 120<br />

3225-0280S 280 125<br />

3225-0300S 300 125<br />

3225-0320S 320 125<br />

3225-0340S 340 125<br />

3225-0360S 360 125<br />

3225-0380S 380 125<br />

3225-0400S 400 125<br />

3225-0420S 420 125<br />

3225-0440S 440 125<br />

3225-0460S 460 125<br />

3230-0280S 280 130<br />

3230-0300S 300 130<br />

3230-0320S 320 130<br />

3230-0340S 340 130<br />

3230-0360S 360 130<br />

3230-0380S 380 130<br />

3230-0400S 400 130<br />

3230-0420S 420 130<br />

3230-0440S 440 130<br />

3230-0460S 460 130<br />

mm fully threaded Locking Screws, TI***<br />

Titanium Diameter Length<br />

REF mm mm<br />

1896-5025S 5.0 25.0<br />

1896-5027S 5.0 27.5<br />

1896-5030S 5.0 30.0<br />

1896-5032S 5.0 32.5<br />

1896-5035S 5.0 35.0<br />

1896-5037S 5.0 37.5<br />

1896-5040S 5.0 40.0<br />

1896-5042S 5.0 42.5<br />

1896-5045S 5.0 45.0<br />

1896-5050S 5.0 50.0<br />

1896-5055S 5.0 55.0<br />

1896-5060S 5.0 60.0<br />

1896-5065S 5.0 65.0<br />

1896-5070S 5.0 70.0<br />

1896-5075S 5.0 75.0<br />

1896-5080S 5.0 80.0<br />

1896-5085S 5.0 85.0<br />

1896-5090S 5.0 90.0<br />

* <strong>Nail</strong>s are packed together with the Set Screw, sterile<br />

** <strong>Long</strong>er Lag Screws are available on request.<br />

*** <strong>Long</strong>er Locking Screws as well as partly threaded screws are available on request.


Ordering Information – Implants<br />

Set Screws, TI (available separately) End Caps, TI<br />

Titanium Diameter Length<br />

REF mm mm<br />

3003-0822S 8.0 17.5<br />

Condyle Screws, TI Nut for Condyle Screw, TI<br />

Titanium Diameter Length<br />

REF mm mm<br />

1895-5040S 5.0 40<br />

1895-5045S 5.0 45<br />

1895-5050S 5.0 50<br />

1895-5055S 5.0 55<br />

1895-5060S 5.0 60<br />

1895-5065S 5.0 65<br />

1895-5070S 5.0 70<br />

1895-5075S 5.0 75<br />

1895-5080S 5.0 80<br />

1895-5085S 5.0 85<br />

1895-5090S 5.0 90<br />

Titanium Diameter Length<br />

REF mm mm<br />

3005-1100S 11.0 0<br />

3005-1105S 15.5 +5<br />

3005-1110S 15.5 10<br />

Titanium Diameter Length<br />

REF mm mm<br />

1895-5001S − 17.0


Ordering Information – Implants<br />

<strong>Long</strong> <strong>Nail</strong> Kit R .0, StSt, Left*, Ø1 . /11mm <strong>Long</strong> <strong>Nail</strong> Kit R .0, StSt, Right*, Ø1 . /11mm<br />

Lag Screw, StSt**<br />

Stainless Steel Length Angle<br />

REF mm °<br />

4320-0280S 280 120<br />

4320-0300S 300 120<br />

4320-0320S 320 120<br />

4320-0340S 340 120<br />

4320-0360S 360 120<br />

4320-0380S 380 120<br />

4320-0400S 400 120<br />

4320-0420S 420 120<br />

4320-0440S 440 120<br />

4320-0460S 460 120<br />

4325-0280S 280 125<br />

4325-0300S 300 125<br />

4325-0320S 320 125<br />

4325-0340S 340 125<br />

4325-0360S 360 125<br />

4325-0380S 380 125<br />

4325-0400S 400 125<br />

4325-0420S 420 125<br />

4325-0440S 440 125<br />

4325-0460S 460 125<br />

4330-0280S 280 130<br />

4330-0300S 300 130<br />

4330-0320S 320 130<br />

4330-0340S 340 130<br />

4330-0360S 360 130<br />

4330-0380S 380 130<br />

4330-0400S 400 130<br />

4330-0420S 420 130<br />

4330-0440S 440 130<br />

4330-0460S 460 130<br />

Stainless Steel Diameter Length<br />

REF mm mm<br />

4060-0070S 10.5 70<br />

4060-0075S 10.5 75<br />

4060-0080S 10.5 80<br />

4060-0085S 10.5 85<br />

4060-0090S 10.5 90<br />

4060-0095S 10.5 95<br />

4060-0100S 10.5 100<br />

4060-0105S 10.5 105<br />

4060-0110S 10.5 110<br />

4060-0115S 10.5 115<br />

4060-0120S 10.5 120<br />

Stainless Steel Length Angle<br />

REF mm °<br />

4220-0280S 280 120<br />

4220-0300S 300 120<br />

4220-0320S 320 120<br />

4220-0340S 340 120<br />

4220-0360S 360 120<br />

4220-0380S 380 120<br />

4220-0400S 400 120<br />

4220-0420S 420 120<br />

4220-0440S 440 120<br />

4220-0460S 460 120<br />

4225-0280S 280 125<br />

4225-0300S 300 125<br />

4225-0320S 320 125<br />

4225-0340S 340 125<br />

4225-0360S 360 125<br />

4225-0380S 380 125<br />

4225-0400S 400 125<br />

4225-0420S 420 125<br />

4225-0440S 440 125<br />

4225-0460S 460 125<br />

4230-0280S 280 130<br />

4230-0300S 300 130<br />

4230-0320S 320 130<br />

4230-0340S 340 130<br />

4230-0360S 360 130<br />

4230-0380S 380 130<br />

4230-0400S 400 130<br />

4230-0420S 420 130<br />

4230-0440S 440 130<br />

4230-0460S 460 130<br />

mm fully threaded Locking Screws, StSt***<br />

Stainless Steel Diameter Length<br />

REF mm mm<br />

1796-5025S 5.0 25.0<br />

1796-5027S 5.0 27.5<br />

1796-5030S 5.0 30.0<br />

1796-5032S 5.0 32.5<br />

1796-5035S 5.0 35.0<br />

1796-5037S 5.0 37.5<br />

1796-5040S 5.0 40.0<br />

1796-5042S 5.0 42.5<br />

1796-5045S 5.0 45.0<br />

1796-5050S 5.0 50.0<br />

1796-5055S 5.0 55.0<br />

1796-5060S 5.0 60.0<br />

1796-5065S 5.0 65.0<br />

1796-5070S 5.0 70.0<br />

1796-5075S 5.0 75.0<br />

1796-5080S 5.0 80.0<br />

1796-5085S 5.0 85.0<br />

1796-5090S 5.0 90.0<br />

* <strong>Nail</strong>s are packed together with the Set Screw, sterile<br />

** <strong>Long</strong>er Lag Screws are available on request.<br />

*** <strong>Long</strong>er Locking Screws as well as partly threaded screws are available on request.


Ordering Information – Implants<br />

Set Screw, StSt (available separately) End Caps, StSt<br />

Stainless Steel Diameter Length<br />

REF mm mm<br />

4003-0822S 8.0 17.5<br />

Condyle Screws, StSt Nut for Condyle Screw, StSt<br />

Stainless Steel Diameter Length<br />

REF mm mm<br />

1795-5040S 5.0 40<br />

1795-5045S 5.0 45<br />

1795-5050S 5.0 50<br />

1795-5055S 5.0 55<br />

1795-5060S 5.0 60<br />

1795-5065S 5.0 65<br />

1795-5070S 5.0 70<br />

1795-5075S 5.0 75<br />

1795-5080S 5.0 80<br />

1795-5085S 5.0 85<br />

1795-5090S 5.0 90<br />

Stainless Steel Diameter Length<br />

REF mm mm<br />

4005-1100S 11.0 0<br />

4005-1105S 15.5 +5<br />

4005-1110S 15.5 10<br />

Stainless Steel Diameter Length<br />

REF mm mm<br />

1795-5001S − 17.0


Ordering Information – Instruments<br />

REF Description<br />

Basic Instruments<br />

702628<br />

1210-6450S<br />

1320-0065<br />

1320-0090<br />

1320-0100<br />

1320-0105<br />

1320-0110<br />

1320-0118<br />

1320-0130<br />

1320-0140<br />

1320-0150<br />

1320-0180<br />

1320-0190<br />

1320-0200<br />

1320-0231<br />

1320-3042S<br />

1806-0041<br />

1806-0095<br />

1806-0096<br />

1806-0185<br />

1320-0215<br />

1806-0232<br />

1320-0315<br />

1806-0325<br />

1806-0365<br />

1806-0480<br />

1320-3042<br />

1806-4270S<br />

1320-9000<br />

1320-6000<br />

T-Handle, Quicklock<br />

Kirschner wire, sterile<br />

Screwdriver 8mm, Ball-Tip, T-handle<br />

<strong>Nail</strong> Holding Screw<br />

<strong>Gamma3</strong> Targeting Arm<br />

Knob for Targeting Sleeve<br />

Clip for K-Wire<br />

Targeting Sleeve 180, green coded<br />

Lag Screw Guide Sleeve<br />

Drill Guide Sleeve 4.2mm for Lag Screw, green<br />

K-Wire Sleeve<br />

Lag Screw Ruler<br />

Lag Screw Step Drill<br />

Lag Screw Driver<br />

Flexible Set Screwdriver, 4mm, small shaft (silicon covered)<br />

Drill Ø4,2 × 300mm, AO small, green, sterile<br />

Awl, Curved<br />

Guide Wire Handle<br />

Guide Wire Handle Chuck<br />

Tissue Protection Sleeve, <strong>Long</strong><br />

<strong>Gamma3</strong> Drill Sleeve<br />

Screwdriver, <strong>Long</strong><br />

<strong>Gamma3</strong> Trocar<br />

Screw Gauge, <strong>Long</strong><br />

Screw Scale, <strong>Long</strong> (for <strong>Long</strong> <strong>Nail</strong>)<br />

Screw Gauge (for <strong>Long</strong> <strong>Nail</strong>)<br />

Drill Ø4,2 × 300mm, AO small, green, unsterile<br />

Drill Ø4,2 × 180mm, AO small, green, sterile (for <strong>Long</strong> <strong>Nail</strong>)<br />

Instrument Tray, Basic, empty<br />

Instrument Set, Basic, completly filled


Ordering Information – Instruments<br />

REF Description<br />

Basic Instruments<br />

Not stored on the Tray<br />

1320-0131<br />

1806-0085S<br />

702634<br />

1320-0112<br />

1320-9002<br />

1806-0032<br />

1320-0210<br />

1320-3042<br />

1320-3045S<br />

Optional Instruments<br />

0152-0218<br />

1320-0041<br />

1320-0042<br />

1213-9091S<br />

1320-0011<br />

1320-0021<br />

1320-0026<br />

1320-0031<br />

1320-0066<br />

1320-0070<br />

1320-0080<br />

1320-0135<br />

1320-0160<br />

1320-0170<br />

1320-3030S<br />

1320-3045S<br />

Lag Screw Guide Sleeve, navigated<br />

Guide Wire, Ball Tip, Ø3 × 1000, Sterile<br />

Large AO Coupling Hall Fitting<br />

<strong>Gamma3</strong> U-Wire<br />

Insert for Bixcut Reamer Heads for Diameter 11, 12, 13, 14, 15.5mm<br />

Trocar for Curved Awl, (Awl Plug)<br />

Straight Screwdriver, 4mm for Set Screw<br />

Drill 4.2 × 300mm, AO small, green, unsterile<br />

Drill 4.2 × 300mm, Tri-Flat Fitting, sterile<br />

K-Wire 1,8×310mm, for Condyle Screws<br />

Cannulated Cutter, use with 4mm Pin only<br />

Sleeve for Cannulated Cutter<br />

Guide Pin 4×400mm, sterile<br />

One Step Conical Reamer,working with Conical Reamer Sleeve short and long<br />

Conical Reamer Trocar, short<br />

Multihole Trocar, short<br />

Conical Reamer Sleeve, short<br />

Spreading Screwdriver<br />

Screwdriver Strike Plate<br />

Universal Joint Socket Wrench<br />

Adaptor for One Shot Device, Gamma<br />

Fragment Control Clip<br />

Fragment Control Sleeve<br />

Drill 3,0×300mm, AO small, sterile, white (for Fragment Control Clip)<br />

Drill 3.0×300mm, Tri-Flat Fitting, sterile, white (for Fragment Control Clip)


Ordering Information – Instruments<br />

0<br />

REF Description<br />

Optional Instruments<br />

1320-3010<br />

1407-4006<br />

1806-0020<br />

1806-0110<br />

1806-0125<br />

1806-0130<br />

1806-0170<br />

1806-0255<br />

1806-0450<br />

1806-0460<br />

1806-4290S<br />

1806-5000S<br />

1320-0022<br />

1320-0027<br />

1320-0032<br />

1320-9002<br />

1320-0210<br />

1320-6010<br />

1320-9400<br />

1320-9005<br />

X-Ray Template<br />

1320-0002<br />

1320-0005<br />

One Shot Device, <strong>Gamma3</strong><br />

<strong>Nail</strong> Extraction Adapter<br />

Guide Wire Ruler (for <strong>Long</strong> <strong>Nail</strong>)<br />

Universal Rod<br />

Reduction Spoon<br />

Wrench, 8mm / 10mm<br />

Slotted Hammer<br />

Condyle Screwdriver (for Condyle Screws)<br />

Tissue Protection Sleeve<br />

Drill Sleeve Ø4.2mm<br />

Drill 4,2×230mm, AO small, sterile, green (<strong>Long</strong> <strong>Nail</strong>)<br />

Drill, 5×230mm, AO small, sterile, black (for Condyle or Shaft Screws)<br />

Conical Reamer Trocar, long<br />

Multihole Trocar, long<br />

Conical Reamer Sleeve, long<br />

Insert for Bixcut Reamer Heads for Diameter 11, 12, 13, 14, 15.5mm<br />

Straight Screwdriver, 4mm for Set Screw<br />

Extraktion Set<br />

Extraktion Tray<br />

Instrument Tray, Optional<br />

X-Ray Template, <strong>Gamma3</strong> <strong>Nail</strong> 180<br />

X-Ray Template, <strong>Gamma3</strong> <strong>Long</strong> <strong>Nail</strong>, R 2.0


References<br />

Publications<br />

More than 1.000.000 Gamma <strong>Nail</strong> implantations have been performed<br />

world wide over the last 17 years. Extensive clinical experience has been<br />

published with the Gamma Locking <strong>Nail</strong>.<br />

We recommend the following publications:<br />

• The Gamma Locking <strong>Nail</strong>, Ten Years Surgical Experience<br />

Gahr, R. H.; Leung, K.-S.; Rosenwasser, M.P.; Roth, W. (eds.),<br />

Einhorn-Presse Verlag, ISBN 3-88756-808-7<br />

• Patients treated with the <strong>Long</strong> Gamma <strong>Nail</strong>,<br />

R. van Doorn,<br />

Bedrijfsnaam: Castellum Drukwerk Vof.<br />

These books contain almost 300 clinical reports available on request.<br />

1


Trauma, Extremities & Deformities<br />

Biologics<br />

Surgical Products<br />

Neuro & ENT<br />

The information presented in this brochure is intended to demonstrate a <strong>Stryker</strong> product. Always refer to the package<br />

insert, product label and/or user instructions before using any <strong>Stryker</strong> product. Surgeons must always rely on their own<br />

clinical judgment when deciding which products and techniques to use with their patients. Products may not be available<br />

in all markets. Product availability is subject to the regulatory or medical practices that govern individual markets.<br />

Please contact your <strong>Stryker</strong> representative if you have questions about the availability of <strong>Stryker</strong> products in your area.<br />

<strong>Stryker</strong> Corporation or its subsidiary owns the registered trademark: <strong>Stryker</strong><br />

<strong>Stryker</strong> Corporation or its subsidiary owns, uses or has applied for the following trademarks: <strong>Gamma3</strong><br />

Literature Number : B 0300009<br />

LOT D3207<br />

Copyright © 2007 <strong>Stryker</strong><br />

Printed in Germany<br />

Trauma, Extremities & Deformities<br />

Biologics<br />

Surgical Products<br />

Neuro & ENT<br />

<strong>Stryker</strong> Trauma GmbH<br />

Prof.-Küntscher-Strasse 1–5<br />

D - 24232 Schönkirchen<br />

Germany<br />

www.osteosynthesis.stryker.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!