Wear performance of Ti based alloy coatings on
316L SS fabricated by sputtering method –
Relevance to Biomedical Implants
G.Godwin
Loyola Institute of Technology & Science
M.Shunmuga Priyan ( iampriyan25@gmail.com )
Loyola Institute of Technology & Science https://orcid.org/0000-0001-8787-6037
S.Julyes Jaisingh
St.Xavier’s Catholic College of Engineering
Research Article
Keywords: Sputtering Method, Roughness, Microhardness, Microabrasion, Adhesion Strength
Posted Date: April 19th, 2022
DOI: https://doi.org/10.21203/rs.3.rs-1565129/v1
License: This work is licensed under a Creative Commons Attribution 4.0 International License.
Read Full License
Page 1/17
Abstract
The present investigations are being carried out to encapsulate 316L SS with a Ti-based alloy coating
fabricated by using PVD sputtering method. TiN, TiO2, and TiCoCr powders with prescribed process
parameters are consecutively coated on 316L SS through PVD sputtering process with coating durations
of 30, 60, and 90 minutes respectively. Further microhardness, surface roughness, microabrasion and
adhesion strength tests were also carried out to determine the impact of the coating material and the
in uence of coating time on the coated surfaces. Comparing to uncoated substrate merely a 60%
improvement in abrasion resistance was observed in TiCoCr coated samples. The XRD results con rmed
the optimal formation of Ti alloy coatings with corresponding orientation over the SS substrates.
Moreover, TiCoCr with a 90 minutes coating duration has much better surface characteristics than TiO2
and TiN. Thus it is preferable that the 90 min coating duration is optimal for coating in steel for bio
implants.
Introduction
The biomaterial is a material that can be implanted to associate with the biological system of a human
body as far as medical devices particularly arti cial joints, hip implants, dental implants, and so on. The
implant biocompatibility plays a key role owing to in vivo environment of humans may give hazardous
products impacting on the tissues and organs (Hallab & Jacobs 2013). Moreover, the medical implant
generally utilizes chrome steel, Co-Cr alloys, and Ti alloys (Sahasrabudhe et al. 2016). For
biocompatibility nature, strength, and invaluable corrosion resistance, stainless steel (SS) is perhaps the
most recognized materials used as biomaterials including iron, chromium, and nickel in loftier in extent,
SS is referred to be a partner especially the iron-based alloys fundamentally will rust or corrode in a
limited capacity to focus (Cao et al. 2014). For biocompatibility nature, strength, and invaluable corrosion
resistance, stainless steel (SS) is perhaps the most recognized materials used as biomaterials including
iron, chromium, and nickel in loftier in extent, SS is referred to be a partner. Martensitic SS are not as
corrosion resistant as ferritic or austenitic SS because of their lower substance of chromium (Holland et
al. 1991). Authors were reported SS for biomaterial implants for the sake of resistance towards human
uids due to the existence of molybdenum (Mo) in the alloy (Sivakumar et al. 1994). Regardless of this
fact, 316L has a downside in that it holds unsatisfactory performance in the wake of tribological behavior
and mechanical strength. This is due to the fact that the hardness value stands low which directed the
experimenters to concentrate on various surface engineering processes so as to enhance its tribological
behavior retaining the distinguishable corrosion resistant property (Sieradzki et al. 1986). With an end
goal to enhance the properties of the biological implant materials such as tribological, electrical, optical,
biological and more, surface coating is employed (Saravanan et al. 2018). As well as that it exhibits
dreadful behavior in wear (Sun and Bell 2002), friction and abrasion (Rahman et al. 2005), microcracks(Li et al. 2019) and severe plastic deformation (Li& Bell 2004). Hence, there is a need for surface
modi cation over the base material 316L SS so as to be utilized as an incomparable biological implant.
(Uddin et al. 2019) hypothesized the tribological, mechanical, and chemical characteristics of TiN
Page 2/17
coatings on Ti substrate for biomedical applications. Authors reported the better adhesion strength,
hardness, wear range, and corrosion rate in TiN coated samples at different parameters (Caha et al.
2019) sprayed TiN coatings using sputtering process. The results showed that TiN coatings deposited
over 80 minutes had superior overall behavior than bare metal and TiN coatings deposited over 30
minutes. The CAE-PVD TiN lm deposited on 316LVM SS and Ti6Al4V alloy was examined by (Lepicka et
al. 2019). Author reported the mechanical characteristics, adhesion, and anti-wear performance, durability
of TiN coated by using PVD deposition technique (Verissimo et al. 2015) and energized ions to be
accelerated to employed higher corrosion resistance, wear resistance (Makhlouf 2011). Microabrasion is
a minimally invasive procedure used to renew overall coating quality and texture. The microabrasive test
helps in simulating the wear conduct of materials under various conditions (Li &Leroux 2016). The
covered and uncoated substrates were permitted to go through wear instrument with the guide of a micro
abrasion tester utilizing a block-on-ring technique, in light of standard ASTM G77 systems as
demonstrated in the schematic representation of micro-abrasive Apparatus (MS Priyan &Hariharan
2014). The geometry of the wear scar, thus formed is proportional to that of the rotating ball, Thereupon
the wear volume can be estimated from either the crater diameter (b) or the depth (t) of the coating
material as follows:
Here, radius of the rotating ball is expressed as R, imposed wear crater depth for the abrasive wear
mechanism as‘t’.
AFM can map out the topography of any surface and represents various physical parameters of the
surface including friction(Gutierrez et al. 2017; Rao and Costa 2014) and to produce the diffraction
pattern X-rays are used because their wavelength, λ is often the same order of magnitude as the spacing,
d, between the crystal planes (1-100 µA) (Pankaew et al. 2012). The goal of this study is aimed to
produce Ti-based alloy coatings on bio implantable material 316L SS utilizing a PVD sputtering
procedure with a exible time frame. Surface features such as abrasion resistance, micro hardness,
coating thickness, and adhesion strength are planned to study evaluate for the specimen. Coating
materials including TiN, TiO2, and TiCoCr are also being considered for usage in the development
process. The coated 316L SS substrates are being studied in order to de ne their coating behavior for
applying the bio implants application.
Experimental Methods
The 316 stainless steel chemical composition mentioned in the Table 1. The commercially available TiN
and TiO2 powder is used for coating process. The proposed novel TiCoCr combination of Ti-55%, Co-25%
Page 3/17
and Cr-20% was mechanically alloyed using high energy planetary ball mill (retsch pm 200, Germany).
Table 1
Chemical composition of 316L SS
Element
C
Mn
Si
P
S
Cr
Mo
Ni
N
Wt. %
0.03
2.0
0.75
0.045
0.03
17
2.5
12
0.1
The substrate material for this experiment is a cylindrical 316 L grade stainless steel with a diameter of
25 mm and a thickness of 1mm, which was cut using the wire-cut EDM (Electrical Discharge Machining)
process. The deposition of Ti based alloy coating target over the base material was performed by sputter
deposition type of physical vapour deposition process. The coating time has been xed for all samples at
30 min, 60 min, and 90min separately. Further, the worn surface morphology of various coating materials
on substrate was analyzed using a scanning electron microscope (Hitachi S-3400, Japan). The crystal
structure analysis of various coating materials on substrate was done using an XRD analyzer (Match
phase analyzer, Germany). The radiation source used was Cu-Kα 1 and the wave length was 1.54 A.
Moreover, the diffracted X-rays were captured using an X-ray lm and thus 2θ peaks were plotted. The
observed 2θ peaks were compared with reference 2θ peaks from [JCDPS] les. Further, the percentage of
each chemical element examined on coated specimen by using energy dispersive X-ray analysis method.
The microabrasion tester using a block-on-ring method, based on standard ASTM G77 procedures as
shown in the schematic representation of micro abrasive apparatus in Fig. 1. The test parameters are
exempli ed in Table 2. The geometry of the wear scar, thus formed was proportional to that of the
rotating ball, there-upon the wear volume was estimated from either the crater diameter (b) or the depth
(t) of the coating material as follows in accordance to Equations.
Here, radius of the rotating ball is expressed as R, imposed wear crater depth for the abrasive wear
mechanism as‘t’. Where, R is the radius of the rotating ball, and‘t’ is the imposed wear crater depth for the
abrasive wear mechanism as shown in Fig. 1.
Page 4/17
Table 2
Micro abrasion test parameters
Parameter
Value
Load range
0.5N, 1N, 1.5N
Diamond ball
High carbon – high chromium material
Sliding distance
150 rev
Ball diameter
25.4 mm
Metal wheel
50 mm
Rubber wheel
50 mm
Abrasive paste
Dry slurry
Motor
0.15/150 kW/rpm
The Vicker’s microhardness measurements were conducted in accordance to ASTM E384. The adhesion
strength between various titanium alloy coating and steel substrate was investigating by the direct pulloff method. In pull-off tests, the coated substrates were glued with epoxy resin of known adhesive
strength to the uncoated steel. The pull-off test was carrying out on an electronic tensile testing machine
FIE UTE-20, India. The test was carried out in compliance with ASTM D5868 guidelines.
Results And Discussion
Figure 2 (a), (b) and (c) shows that the XRD results of coated TiCoCr, TiN and TiO2 on 316L SS
specimens. The samples were analyzed over a range from 10° to 80°. Figure 2 (a) shows the sharp peaks
and proves the formation of Titanium at 35.93, 40.142 and 53.01, Cobalt at 62.925, 75.867 and
Chromium at 44.244, 38.386. The crystal plane index was found for Titanium, Cobalt and Chromium
such as (101), (110) and (100) and it was evident in JCPDS le number 653362, 897373, 892871,
respectively. Owing to their minimum weight percentage proportion, the intensity of diffraction peaks was
low for Co and Cr corresponds with titanium. Similarly, Fig. 2 (b) shows the characterization of sharp
peaks and proves the formation of TiN located at 36.675, 42.797, 61.857 and 74.445. The crystal plane
index was found for TiN such as (111), (200), (220), (311) and it was evident in JCPDS le number
870633. Similar indications were noted in Fig. 2 (c) the XRD graph of TiO2 coated surface on 316L SS.
The strong 2 theta peaks at 25.102, 36.485 and 48.912 equal to the miller indices plane (101), (004) and
(200) were con rms the presence of Ti and O molecules in the coated surface. Thus the coating on 316L
SS is effective, which produce considerable amount of dopants on the substrate. Figure 3 (a), (b) and (c)
shows the EDAX spectrum of TiCoCr, TiN and TiO2 coated 316L steel specimens. It is noted that the
spectrum for TiCoCr gives the atoms of Co, Cr and Ti with different keV values.
Page 5/17
Table 3 shows the micro abrasion properties such as wear volume and speci c wear rate of various
coating done on stainless steel substrate. The incremental in wear volume of 71% and 261.8% was noted
for 1 N and 1.5N loading conditions. This notably higher wear volume and speci c wear rate was the
reason for adhesion and three body abrasion wear loss mechanism. Thus higher wear loss is evidenced
at higher loading condition. Figure 4 shows the SEM worn surface tracks of uncoated 316L SS
specimens at different loading conditions.
Table 3
Wear behavior of uncoated 316L SS substrate
Coating time ( min)
Uncoated
Load
Wear volume
Speci c wear rate
(N)
×10− 5 m3
×10− 14 mm3/Nm
0.5
18.54
1.027
1
31.72
0.881
1.5
67.08
0.931
Table 4 shows the wear volume and speci c wear rate of TiN coated 316L SS substrate with different
loading conditions and coating time. It is noted that the wear volume of 16.5, 12.8 and 10x10− 5 mm3 was
observe for 1.5N load at 30, 60 and 90 min coating time respectively. Table 5 shows the TiO2 coating on
316L SS at 0.5, 1.0 and 1.5N loads and 30, 60 and 90 min coating time. The wear volume and wear rate
is signi cantly higher for TiO2 coating by 30 and 60 min. This increment in the wear volume and speci c
wear rate is the reason for no strengthening mechanism of TiO2 on steel surface. Moreover, there is no
evidence for pitting marks and dimple, which con rms the high wear resistance behavior of material even
after the shear force is applied.
Table 6 shows the wear volume and war rate is signi cantly reduced for TiCoCr coating by 30, 60 and 90
min. On comparing to the uncoated surface the TiCoCr coated steel gives very high wear resistance. A
very lowest wear volume of 7.372×10− 5m3 is observed for coating done for 90 min at 0.5N loading
condition. Moreover in 90 min coating time all the loading values gives improved abrasion resistance.
This is because of higher penetration time of Co and Cr atoms onto the surface of steel. The addition of
Cr and Co may produce ne intermetallic on the substrate’s surface, thus giving higher wear resistance.
Both the Co and Cr may reinforce into the surfaces of steel and produced high surface rigidity and also
giving lesser 3 body abraded particle and wavy structure.
Page 6/17
Table 4
Wear behavior of TiN coated 316L SS substrate
Coating time ( min)
30
60
90
Load (N)
Wear volume
Speci c wear rate
×10− 5 m3
×10− 14 mm3/Nm
0.5
15.9
0.883
1
16.3
0.452
1.5
16.5
0.229
0.5
9.7
0.538
1
9.9
0.275
1.5
12.8
0.177
0.5
9.4
0.522
1
9.5
0.263
1.5
10
0.138
Table 5
Wear behavior of TiO2 coated 316L SS substrate
Coating time ( min)
30
60
90
Load (N)
Wear volume
Speci c wear rate
×10− 5 m3
×10− 14 mm3/Nm
0.5
21.94
1.219
1
22.79
0.633
1.5
23.38
0.324
0.5
18.84
1.047
1
20.25
0.562
1.5
20.94
0.290
0.5
13.51
0.750
1
13.66
0.379
1.5
14.15
0.196
Page 7/17
Table 6
Wear behavior of TiCoCr coated 316L SS substrates with different load
and coating time
Coating time ( min)
30
60
90
Load (N)
Wear volume
Speci c wear rate
×10− 5m3
×10− 14 mm3/Nm
0.5
13.017
0.722
1
13.360
0.371
1.5
13.54
0.188
0.5
8.212
0.456
1
8.405
0.233
1.5
9.074
0.126
0.5
7.372
0.409
1
7.524
0.208
1.5
7.600
0.105
Figure 5 (a), (b) and (c) observed COF of 0.09, 0.053, 0.05 and 0.041 in 0.5 N load, 0.118, 0.082, 0.091
and 0.077 in 1N and 0.311, 0.169, 0.178 and 0.123 in 1.5N were noted for coating time 30 min at different
loads. At higher loading condition such as 1.5N the uncoated 316L SS substrate gives higher COF. This is
because of higher a nity of uncoated surface to the abrasion ball and lesser sliding velocity. On
comparing to all coating time the 90 min coating time gives typically lesser COF, which indicates the
higher wear resistance of 90 min coated substrates.
Figure 6 shows the micro hardness values of coated surfaces with respect to the loads. It is noted that
the uncoated surfaces have the surface hardness of 210HV at 5N, 191HV at 10N, 179HV at 20N, 175HV
at 30N and 172HV at 40N. This nominally lesser hardness is the reason for coarse grain structure in the
substrate material. However the doping of reinforcement such as TiN, TiO2 and TiCoCr onto the 316L SS
surface the surface hardness is increased. This is because of adding Ti atoms to the surface along with
N, Co and Cr. Figure 7 shows the AFM resulting of TiN, TiO2 and TiCoCr coated on the 316L SS substrate
at 30, 60 and 90 min respectively. Moreover, the coated grains were coarser and atter, which increased
the plastic deformation rate. The AFM image of TiCoCr coated 316L SS at 30, 60 and 90 min coating
time. The thickness of 110nm, 170nm and 260nm were achieved, which is eventually higher than TiN and
TiO2 coatings on the substrate.
Figure 8 and 9 shows the graph of surface roughness (Ra) and the adhesion strength of various coating
done using 316L SS substrate with different coating time. It is observed that the surface roughness of 30,
60 and 90 min coated surfaces show signi cant changes in the surface roughness. The coated surfaces
at the lesser coating time (30 min), the surface roughness is typically higher and the same decreases with
Page 8/17
increase of coating time up to 90 min. It is noted that the surface roughness of 0.31, 0.38 and 0.28µm
were noted for TiN, TiO2 and TiCoCr coating on 316L SS substrate at the lesser coating time 30 min.
However the aged coating time up to 60 and 90 min led lesser surface roughness of 0.25, 0.33, 0.22 and
0.19, 0.26, 0.17µm for TiN, TiO2 and TiCoCr coating at 60 and 90 min respectively. The values showed
reduction trend in surface roughness with respect to the coating time. This reduction in the coating
roughness is the reason for uniform lling of atom at the aged coating process.
Figure 9 shows the adhesion strength of various coatings on 316L SS substrate with various coating
time. It is noted that the 30 min coating time in TiO2, TiN and TiCoCr gives marginal adhesion strength of
26.7, 32.4 and 34.2 MPa. However when the coating time increases the there is a marginal shift in
adhesion strength. The coating time 60 min yield improved adhesion strength of 28.8, 35.5 and 36.4MPa,
which is equal to 7.86%, 9.56% and 6.43% on compare to the 30 min coating time. This improvement is
the reason for effective layer building and high cohesive strength of particle. When the coating time is
higher the particle settling and the cohesive force between the TiO2, TiN and TiCoCr increases, which
resulted higher adhesion strength when the shear force is applied. It is further noted that the prolonged
coating time up to 90 min signi cantly reduced the adhesion strength. There is a marginal dip in the
strength is observed in all coating materials such as TiO2, TiN and TiCoCr respectively. The observed
adhesion strength of 28.1, 34.8 and 35.8MPa were noted for 90 min coating time in TiO2, TiN and TiCoCr
respectively. This is about 2.43%, 1.97% and 1.64% of decrement on compare with the 60 min coating on
the substrate.
It is noted that on comparing TiN, TiO2 and TiCoCr the higher adhesion strength of 36.4MPa was
observed for 60 min coated TiCoCr on 316L SS substrate. This higher adhesion strength for TiCoCr is the
reason for higher cohesive force of atoms with each other than the 90 min coated substrate. Moreover it
is noted that by comparing all the TiCoCr coating gave improved results in 60 min coating time. This is
the reason for high cohesion strength of Ti, Co and Cr atoms and their penetration ability. Thus the 60
min coating time yields higher adhesion strength in all coated substrates.
Conclusions
The speci c conclusions made from this present investigation are as follows.
1. The sputtering process successfully yields signi cant coating thickness over the stipulated coating
time. The results of coating were con rmed using XRD and EDAX analysis.
2. The titanium alloy TiCoCr coating at 90 min on 316L SS gave higher abrasion resistance among
other. Approximately 60% improvement was observed than that of uncoated substrate. A lesser
speci c wear rate of 0.105×10–14 mm3/Nm with COF of 0.023. However the TiN and TiO2 coating on
substrate yield lower value.
3. The microhardness of TiCoCr 90 min coated 316L SS implants gave highest Vicker’s hardness
number. The Vicker’s hardness number of 682 HV was observed for 90 min coated TiCoCr 316L SS
Page 9/17
implant. The other coating time yields lesser microhardness.
4. The surface roughness of 90 min coating time on 316L SS substrate with TiO2, TiN and TiCoCr gave
highly smooth surfaces whereas at 30 and 60 min the surface roughness marked higher.
5. The adhesion strength of 60 min coated samples yields improved results. Highest adhesion strength
of 36.2MPa was observed for 60 min coated TiCoCr on 316L SS implant. However the 30 and 90 min
coatings gave marginal dip on the adhesions strength.
. Thus according to the ndings the titanium alloy TiCoCr with 90 min coating time gave much
improved surface properties than oxide state of TiO2 and nitrate state of TiN. Thus it is preferable
that the 90 min coating time is idle for coating titanium based alloys in steel based bio implants.
Moreover instead of the oxide and nitrate state of titanium the alloy state of TiCoCr could be
preferable in order to achieve desirable results on the bio implants.
Declarations
The authors declare no competing interests
References
1. Hallab, N.J and Jacobs J.J, Orthopedic Applications in Biomaterials Science: An Introdcution to
Materials, Third Edition, (2013), pp 841–882.
2. Sahasrabudhe, H and Soderlind, J & Bandyopadhyay, A, Laser processing of an insitu TiN/Ti
composite coating on Titanium, Journal of the Mechanical Behaviour of Bio Materials, Vol. 53,
(2016), pp. 239–249.
3. Cao, Y, Wang, Y, An, X, Liao, X, Kawasaki, M & Ringer, S, ‘Concurrent microstructural evolution of
ferrite and austenite in a duplex stainless steel processed by high-pressure torsion’, Acta Materialia,
(2014), vol. 63, pp. 16–29.
4. Holland Roy, I, ‘Use of potentiodynamic polarization technique for corrosion testing of dental alloys’,
European Journal of Oral Sciences, (1991) vol. 99, no. 1, pp. 75–85.
5. Sivakumar M, Mudali UK, Rajeswari S, In vitro electrochemical investigations of advanced stainless
steels for applications as orthopedic implants, Journal of Materials Engineering Perform, vol. 3,
(1994), pp 744–753.
. Sieradzki, K & Newman, R ‘A Percolation Model for Passivation in Stainless Steels’, Journal of the
Electrochemical Society, vol. 133, no. 9 (1986) pp. 1979–1980.
7. Saravanan, M, Devaraju, A, Venkateshwaran, N, Krishnakumari,, A and Saavesh, J, A review on recent
progress in coatings on AISI austenitic stainless steel, Materials Today: Proceedings, vol. 5, issue no
6 (2018) pp. 14392–14396.
. Sun Y, & Bell, T, Dry sliding wear resistance of low temperature plasma carburized austenitic stainless
steel, Wear, vol. 253, issue no. 5–6 (2002) pp.689–693.
Page 10/17
9. Rahman, M, Haider, J & hashmi, M.S.J, Low temperature plasma nitirding of 316 stainless steel by a
saddle eld fast atom beam source, Surface Coating Technology, vol. 200, issue no 5–6 (2005)
pp. 1645–1651.
10. Li, B, Zheng, Y, shi, S, Liu, Y & Chen, X, Microcrack initiation mechanism of 316 LN austenitic
stainless steel under in-phase thermo mechanical fatigue loading, Material Science and Engineering:
A, vol.752, (2019) pp 1–14.
11. Li, C.X and Bell, T, Sliding wear properties of active screen plasma nitride 316 austenitic stainless
steel, Wear, vol. 256, issue no. 11–12 (2004) pp 1144–1152.
12. Uddin, G, Jawad, M, Ghufran,M, Saleem,M, Raza, M and Rehman, Z, ‘Experimental investigation of
tribolo-mechanical and chemical properties of TiN PVD coating on titanium substrate for biomedical
implants manufacturing’, The International Journal of Advanced Manufacturing Technology,
vol. 102, issue no. 5–8 (2019) pp. 1391–1404.
13. Çaha, I, Alves, A, Affonço, L, Lisboa-Filho, P, da Silva, J & Rocha, L, ‘Corrosion and tribocorrosion
behaviour of titanium nitride thin lms grown on titanium under different deposition times’, Surface
and Coatings Technology, vol. 374 (2019) pp. 878–888.
14. Lepicka, M, Gradzka-Dahlke, M, Pieniak, D, Paierbiewicz, K, Krynska, K and Niewczas, A. ‘Tribological
performance of titanium nitride coatings: A comparative study on TiN-coated stainless steel and
titanium alloy’, Wear, vol. 422–423 (2019) pp. 68–80.
15. Verissimo, N.C, Chung, S and Webster, T.J. New nonoscale surface modi cations of metallic
biomaterials. In Surface Coating and Modi cation of metallic Biomaterials, (2015) pp. 249–273.
1 . Makhlouf, A.S.H. Current and Advanced coating technologies for industrial applications. In
Nanocoatings and Ultra-Thin Films, pp. 3–23.
17. Li, D, lerpoux P. Block on ring sliding wear evaluation: (2016)
1 . Priyan MS, Hariharan P, Wear and corrosion resistance of Fe based coatings by HVOF sprayed on
gray cast iron for automotive application, Tribology in industry, vol. 36 (2014) pp. 394–405.
19. Gutierrez, R. M.P, Mendez, J.V, and Vazquez, I. A, A novel approach to the oral delivery of bio
nanostructures for systemic disease, In Nano structures for Oral Medicine (2017) pp. 27–59.
20. Pankaew, P, Hoonnivathana, E, Sujinnapram, S, Thamphat, K, Limsuwan, P and Naemchanthara, K,
‘Characterization of Apatitte from Human Teeth via XRD, Ft = IR and TGA Techniques’, Advanced
Materials Research, vol.506 (2012) pp. 90–93.
Figures
Page 11/17
Figure 1
Schematic diagram of abrasion tester and its Mechanism
Page 12/17
Figure 2
XRD graph of (a) TiCoCr (b) TiN (c) TiO2 coating on 316L SS
Figure 3
Page 13/17
SEM fractograph with EDX of (a) TiCoCr (b) TiN (c) TiO2 -316L SS substrate
Figure 4
SEM worn surface tracks of Ti based coated at different loading conditions
Page 14/17
Figure 5
COF of (a) 30 min (b) 60 min (c) 90 min coated substrates
Figure 6
Page 15/17
Micro hardness values of coated and uncoated 316L SS substrate
Figure 7
AFM images of Ti based coating on 316L SS at (a) 30 min (b) 60 min (c) 90 min
Figure 8
Page 16/17
Surface roughness of coatings on substrate
Figure 9
Adhesion strength of coatings on substrate
Page 17/17